Analog Output Calibration 

To perform the analog output calibration, it is 
recommended to use a milliamp meter with a 
resolution of at least 0.1 µA to measure the output 
current. The values saved internally during this 
procedure are used for scaling the 4-20 mA output in 
the 

Setup

 menu. 

 

Analog Output Calibration 
Procedure 

1.  Wire the PD7000 4-20 mA output to a current 

loop that includes a power supply (internal or 
external 12 to 24 VDC), and the mA input on the 
digital meter. See 

Figure 20. 4-20 mA Output 

Connections

 on page 

15

 for details. 

2.  Turn on all devices. Allow for a 15 to 30 minute 

warm-up. 

3.  Go to the Advanced Features menu, and 

navigate to the Analog Output Programming 
(

AoutPr

)/Calibration (

Calib

) menu and press 

Enter

.  

4.  The display will show 

4 nmA

. The PD7000 mA 

output should now be close to 4 mA. Press Enter 
and the display will show 04.000. Enter the actual 
value read by the digital mA meter and press 

Enter

5.  The display will show 

20 nmA

. The PD7000 mA 

output should now be close to 20 mA. Press 

Enter

 and the display will show 20.000. Enter the 

actual value read by the digital mA meter and 
press 

Enter

6.  The meter will now calculate the calibration 

factors and store them.  

7.  Press 

Menu

 to exit and return to Run mode. 

 

Programmable Function Keys 
User Menu (

user

The 

User 

menu allows the user to assign the front 

panel function keys F1, F2, and F3, the digital input 
F4 (a digital input located on the input signal 
connector), and up to eight additional digital inputs to 
access most of the menus or to activate certain 
functions immediately (e.g. reset max & min, hold 
relay states, etc.). This allows the meter to be greatly 
customized for use in specialized applications. 

Up to eight digital outputs can be assigned to a 
number of actions and functions executed by the 
meter (i.e. alarms, relay acknowledgement, reset 
max, min, or max & min, tare, and reset tare). The 
digital outputs can be used to trigger external alarms 
or lights to indicate these specific events. 
 

 

 

Function Keys & Digital I/O 
Available Settings 

Refer to the following table for descriptions of each 
available function key or digital I/O setting. 

Display 

Description 

Rst Hi 

Reset the stored maximum display value 

Rst Lo 

Reset the stored minimum display value 

Rst HL 

Reset the stored maximum & minimum 
display values 

tare 

Capture tare and zero the display 

Rst tr 

Reset captured tare and resume normal 
operation 

relay 

Directly access the relay menu 

Set 1

*

 

Directly access the set point menu for relay 1 
(*through 8) 

Rly d 

Disable all relays until a button assigned to 

enable relays

 (

Rly E

) is pressed 

Rly E 

Enable all relays to function as they have 
been programmed 

O Hold 

Hold current relay states as they are until a 
button assigned to 

enable relays

 (

Rly E

) is 

pressed 

d Hold 

Hold the current display value momentarily 
while the function key or digital input is 
active. The process value will continue to be 
calculated in the background. 

Ln1 Hi 

Display maximum display value on line 1  

Ln1 Lo 

Display minimum display value on line 1 

Ln1 HL 

Display maximum & minimum display values 
on line 1 

Ln2 Hi 

Display maximum display value on line 2 

Ln2 Lo 

Display minimum display value on line 2

 

Ln2 HL 

Display maximum & minimum display values 
on line 2 

F On 1

*

 

Force relay 1 (*through 4) into the on state. 
This function is used in conjunction with a 
digital input expansion module to achieve 
interlock functionality. See page 27 for 
details about interlock relays. 

Contrl 

Directly access the control menu 

disabl 

Disable the selected function key or digital 
I/O 

AcH 

Acknowledge all active relays that are in a 
manual operation mode such as auto-
manual or latching 

Reset 

Directly access the reset menu 

nmenu 

Mimic the menu button functionality (digital 
inputs only) 

rigHt 

Mimic the right arrow/F1 button functionality 
(digital inputs only) 

up 

Mimic the up arrow/F2 button functionality 
(digital inputs only) 

Enter 

Mimic the enter/F3 button functionality 
(digital inputs only) 

Alnm 1

*

 

Provide indication when alarm 1 (*through 8) 
has been triggered (digital outputs only) 

 

*If installed

 

Содержание ProVu PD7000

Страница 1: ...with Interlocking Capability Isolated 4 20 mA Output Options Free PC Based Onboard MeterView Pro USB Programming Software No Assembly Required Optional SunBright Display Models for Outdoor Applicatio...

Страница 2: ...gistered Trademarks All trademarks mentioned in this document are the property of their respective owners 2020 Precision Digital Corporation All rights reserved FREE MeterView Pro Programming Software...

Страница 3: ...tions 15 Analog Output Power Supply 15 External Relays Digital I O Connections 16 Interlock Relay Feature 16 Setup and Programming 17 Front Panel Buttons and Status LED Indicators 17 Display Functions...

Страница 4: ...Making Changes to a Password Protected Meter 29 Disabling Password Protection 29 Advanced Features Menu 30 Advanced Features Menu Display Messages 30 Offset Adjust Adjust 31 Recalibration of the Mete...

Страница 5: ...1 Thermocouple Input Connections 14 Figure 12 Three Wire RTD Input Connections 14 Figure 13 Two Wire RTD Input Connections 14 Figure 14 Four Wire RTD Input Connections 14 Figure 15 Average Temperature...

Страница 6: ...to any system Ordering Information Standard Models 85 265 VAC Model 12 24 VDC Model Options Installed PD7000 6R0 PD7000 7R0 No options PD7000 6R2 PD7000 7R2 2 relays PD7000 6R3 PD7000 7R3 4 20 mA outp...

Страница 7: ...igital inputs Pass 2 Allows use of function keys digital inputs and editing set reset points Pass 3 Restricts all programming function keys and digital inputs Non Volatile Memory All programmed settin...

Страница 8: ...50 0 4 C 120 110 to 500 0 2 F 79 to 260 0 1 C 1000 328 to 900 0 7 F 200 to 482 0 4 C All ranges capable of 1 or 0 1 resolution Relays Rating 2 or 4 SPDT Form C internal and or 4 SPST Form A external r...

Страница 9: ...wer USB port provides power to the meter DO NOT apply AC or DC power to the meter while the USB port is in use On Board Digital Input F4 Function Remote operation of front panel buttons acknowledge re...

Страница 10: ...10V 0 15 80 MHz 1kHz 80 AM AC Surge 2kV Common 1kV Differential Surge 1KV CM Power Frequency Magnetic Field 30 A m 70 V for 0 5 period Voltage Dips 40 V for 5 50 periods 70 V for 25 periods Voltage I...

Страница 11: ...gh the bracket Slide the bracket toward the front of the case and remove Insert meter into the panel cutout Install mounting brackets and tighten the screws against the panel To achieve a proper seal...

Страница 12: ...at www predig com meterviewpro MeterView Pro Installation 1 Connect one end of the provided USB cable to the meter and the other end to the computer The computer will automatically install the driver...

Страница 13: ...ent and the meter Power Connections Power connections are made to a two terminal connector labeled POWER The meter will operate regardless of DC polarity connection The and symbols are only a suggeste...

Страница 14: ...ion provides sufficient lead wire compensation to obtain accurate readings even with long leads Figure 11 Thermocouple Input Connections Figure 12 Three Wire RTD Input Connections Figure 13 Two Wire R...

Страница 15: ...DC Loads Protection RC Networks Snubbers Available from Precision Digital RC networks are available from Precision Digital and should be applied to each relay contact switching an inductive load Part...

Страница 16: ...s the interlock relay feature reassigns one or more alarm control relays for use as interlock relay s Interlock contact s are wired to digital input s and activate the interlock relay This feature is...

Страница 17: ...eter Front Panel Buttons and Status LED Indicators Button Symbol Description LED Status Menu 1 8 Alarm 1 8 indicator Right arrow F1 1 8 M Flashing Relay in manual control mode Up arrow F2 M Manual con...

Страница 18: ...nual reset only after alarm condition has cleared Altern Alternate Set relay for alternation control Sanmpl Sampling Set relay for sampling operation Display Functions Messages Display Parameter Actio...

Страница 19: ...3 Set or enter Password 3 unloc Unlocked Program password to lock meter locd Locked Enter password to unlock meter 99999 9999 Flashing display Overrange condition Underrange condition Open Open Open s...

Страница 20: ...Up arrow button to scroll through choices Press the Menu button to exit at any time Setting the Input Signal Input Enter the Input menu to set up the meter to accept thermocouple tc or RTD rtd inputs...

Страница 21: ...to give the best performance under various lighting conditions Select intensity 8 for outdoor applications The default intensity is 6 Setting the Display Units or Custom Tags units Enter the display...

Страница 22: ...feature enabled 2 Set point 3 Reset point 4 Fail safe operation a On enabled b Off disabled 5 Time delay a On delay 0 999 9 seconds b Off delay 0 999 9 seconds 6 Relay action for loss break of input...

Страница 23: ...nd Off time delays may be programmed for each relay between 0 and 999 9 seconds The relays will transfer only after the condition has been maintained for the corresponding time delay The On time delay...

Страница 24: ...point the relay trips and the sample time starts After the sample time has elapsed the relay resets The cycle repeats every time the set point is crossed going up for high alarms and going down for lo...

Страница 25: ...gh or low temperature alarms control applications such as simple on off temperature control There are four basic ways the relays can be used 1 High or Low Alarms with Latching or Non Latching Relays 2...

Страница 26: ...n Relay terminology for following tables Terminology Relay Condition On Alarm Tripped Off Normal Reset Ack Acknowledged The On and Off terminology does not refer to the status of the relay s coil whic...

Страница 27: ...the ACK button or the assigned digital input is closed all relays programmed for manual reset are acknowledged Setting Up the Interlock Relay Force On Feature Relays 1 4 can be set up as interlock re...

Страница 28: ...ed with a decimal point regardless of the decimal point selection for the temperature display For instructions on how to program numeric values see Setting Numeric Values page 19 Reset Menu reset The...

Страница 29: ...ress the Enter button while the message is being displayed and enter the correct password to gain access to the menu Note After exiting the programming mode the meter returns to its password protected...

Страница 30: ...Action Setting Parity Parity Select parity Even Odd or None with 1 or 2 stop bits t byt Time byte Set byte to byte timeout SelEct Select Enter the Select menu AoutPr Analog output programmin g Program...

Страница 31: ...of accuracy to the calibration The scale can be changed to the Celsius scale after calibration is completed The meter will display temperature accurately in any scale The following table shows the re...

Страница 32: ...ion Digital s expansion modules cables or meters to the RJ45 M LINK connector Otherwise damage will occur to the equipment and the meter Note More detailed instructions are provided with each optional...

Страница 33: ...n be used to trigger external alarms or lights to indicate these specific events Function Keys Digital I O Available Settings Refer to the following table for descriptions of each available function k...

Страница 34: ...reference enter the room temperature e g 25 0 and press Enter 7 The meter displays 0mv apply an input equal to 0 000 mV and press Enter The display flashes while accepting the 0 mV input press Enter...

Страница 35: ...default settings for F1 F2 and F3 F4 Operation A digital input F4 is standard on the meter This digital input is programmed identically to function keys F1 F2 and F3 The input is triggered with a con...

Страница 36: ...see Advanced Features Menu Display Messages on page 30 Determining Software Version To determine the software version of a meter 1 Go to the Diagnostics menu diAG and press Enter button 2 Press Up ar...

Страница 37: ...relay 2 Off 2 0 0 sec On delay relay 3 On 3 0 0 sec Off delay relay 3 Off 3 0 0 sec On delay relay 4 On 4 0 0 sec Parameter Display Default Setting Off delay relay 4 Off 4 0 0 sec Input break relay 1...

Страница 38: ...onse is too slow Check filter and bypass values Display reading is not accurate Check 1 TC wire type 2 Input selection in Setup menu Display does not respond to input changes reading a fixed number Ch...

Страница 39: ...le technical standards and when applicable or required a European Union notified body certification Standards EN 55022 2003 EN 61000 6 2 2001 EN 61010 1 2001 EN 61326 2006 The standards EN 55022 2003...

Страница 40: ...predig com LIM7000_I SFT063 Ver 4 000 up 05 20 Contact Precision Digital Technical Support Call 800 610 5239 or 508 655 7300 Fax 508 655 8990 Email support predig com Sales Support Call 800 343 1001 o...

Отзывы: