9 Working with the configuration menu
■
21
En
gl
is
h
The value set for „FLOATINGDISPLAY“ defines the period after which each new measurement is displayed.
for the definition of this period, the quality of the balance location is crucial. The stability control must also
set appropriately.
Recommended values:
– Optimum balance location:
„FLOATINGDISPLAY 0.08“
– Good balance location:
„FLOATINGDISPLAY 0.16“
– Critical balance location:
„FLOATINGDISPLAY 0.32“
The value set for the „STABILITY“ control depends on the quality of the balance location and must be cor-
rectly chosen in order to obtain optimum, reproducible results. Choose:
– Optimum balance location:
„STABILITY LOW“
– Good balance location:
„STABILITY MEDIUM“
– Critical balance location:
„STABILITY HIGH“
With the aid of the „AUTO-STANDBY“ function, you can define the period of non-use before the balance au-
tomatically switches into the energy-saving mode.
The Auto-Standby mode turns off the balance automatically, if:
• the balance is tared and has shown “Zero” for at least 5 minutes
• the balance has received no remote control command via the interface for at least 5 minutes,
• the automatic zero correction „AUTO-ZERO” is activated.
It is possible to re-start the balance after it has been switched off by an automatic Auto-Standby:
• Briefly press any key
• Put a weight on the balance
• Make a remote control command via the interface
If the automatic zero correction „AUTO-ZERO” is activated, the balance always gives a stable zero (e.g. even
with fluctuations in room temperature).
The „QUICK-TARE“ forces the balance to perform a tare immediate when pressing «
0/T
» or sending the re-
mote command to tare, independent of the stability of the balance.
The „ZERO-RANGE“ defines the maximim weigth where the balance will performe a zero function instead of
a normal tare when pressing «
0/T
» or sending the remote command to tare. This value can be adjusted to
make sure a small load is really saved in the tare and therefore printout e.g. on a report as tare value.
FLOATINGDISPLAY
0.08
0.16
0.32
Input integration time (in seconds)
STABILITY
LOW
MEDIUM
HIGH
Setting the stability control (instability of the
balance location)
AUTO-STANDBY
OFF
0.5 MIN.
1 MIN.
5 MIN.
10 MIN.
Auto-Standby not active or active after nn min-
utes
AUTO-ZERO
ON
/OFF
Automatic zero correction on/off
QUICK-TARE
ON/
OFF
Quick tare on/off
ZERO-RANGE
0.000
g
Maximum weight where a zero function in per-
formed
NOTE
The value of the Floating Display is a function of the stability control and the balance location. For balance
location, see chapter 7.4 "Choosing a suitable location" and see chapter 9.6 "Weighing mode".
NOTE
The Auto-Standby function only works with the automatic zero-correction activated („AUTO-ZERO“).