11 Statistics
30
11
Statistics
See Chapter 8.5 “Activating the two main menus” and Chapter 8.6 “How the menu control operates” for the
activation of the application menu.
With „MODE“ you may define whether only the statistics-program, only the storage-program or both pro-
grams simultaneously should be used.
The „COUNT“ defines the number of measurements is laid down, after which automatic storage is to be ter-
minated.
With „RECORDING“ you can choose between „MANUAL“ storing a weight value by pressing the «
REF
»-but-
ton, storing the value afer each „LOADCHANGE“ or storing every value measured after a defined period with
„TIMEBASE“.
The „TIMEBASE“ defeines the time span for the recording of data in accordance with „RECORDING TIME-
BASE” (for example, every 2.0 seconds)
.
Functions of the application:
- - 0000
The balance carries out a Zero measurement again.
measurement 2
EXIt
- - iNT / LLLL
The internal weight is applied and measured again or place the same ex-
ternal weight again.
measurement 1
EXIt
The balance repeates this measurements for 10 cycles.
+0,00008
9
The standard deviation of the measurement is calculated and displayed
and a report is printed.
STANDARD-DEV.
exit
Press «
REF
»to exit the repeatability test at anytime.
• SET STATISTIC
STATISTIC
MODE
OFF
STATISTIC
RECORDER
STAT./RECORDER
Statistics program off
Statistics only
Data storage only
Statistics and storage
COUNT
100
Number of values to be stored automatically
(1..500).
RECORDING
MANUAL
TIMEBASE
LOADCHANGE
With
«STO i»
function key
on a time basis
after every weight change
TIMEBASE
2.0
Time base for “storage” in seconds
NOTE
In storing the first value a range of ±50% is determined. Subsequent values must be within this range
otherwise an error message will be issued.
«STO i»:
- Take value, start/stop of automatic recording
- Short acustic confirmation
«AUTO i»:
- Automatic recording is running
«WAIT i»:
- Value not stabilized yet
«RES»:
- Before a new series of measurements the storage must be reset using «
RES
».
An
acoustic
signal sounds and the measurement counter is reset.
«END»:
- Permanently store data (only with recording activated
Display
Step