9 Working with the configuration menu
20
With “SET PRINTFORMAT”, elements which are switched on are printed in each case:
• with “UNITS” all momentarily active units are printed out,
• with “LINEFEED” the specified numbers of lines are printed out or a formfeed is performed the end of each
printout,
• with “PRODUCT ttt...” the product name can be entered alphanumerically,
• with “PRODUCTMODE HOLD” this product name is stored,
• with “PRODUCTMODE DELETE” it is deleted after each expression,
• with “PRODUCTMODE COUNT” a counter, which is incremented by 1 after each expression,
• “OPERATOR ttt...” is the alphanumeric input of the operator.
When a peripheral device (for example a printer) is connected, the balance interface must be configured in
the submenu “SET INTERFACE” (see Chapter 9.7 “Interface-functions”).
9.5
Calibration functions
To activate the menu see Chapter 8.5 “Activating the two main menus” and Chapter 8.6 “How the menu
control operates”.
For the calibration of the balance see Chapter 7.7 “Calibration of the balance” and Chapter 16.1 “Calibrati-
on”.
9.6
Weighing mode
To activate the menu see Chapter 8.5 “Activating the two main menus” and Chapter 8.6 “How the menu
control operates”.
PRODUCTMODE
HOLD
DELETE
COUNT
OPERATORttt...
• SET
CALIBRATION
MODE OFF
Closed
EXTERNAL
External
EXT.-DEF.
External with user-defined weight („DEF. n.nnn
g“)
INTERNAL
with internal weight
AUTO
Automatic (AUTOCAL)
DEF.
0.000
g
Calibration weight for EXT.-DEF. mode
AUTOCAL.
TIME/TEMP.
TEMPERATURE
TIME
Auto-calibration on time and temperature
Auto-calibration on temperature
Auto-calibration on time
AUTOCAL.-TIME
6 h
Time for auto-calibration
AUTO-WARNING
ON/
OFF
Shows a message if the balance recognizes a si-
gnificant temperature change and therefore a
calibration is recommended
REPORT
ON
/OFF
Printout calibration report
NOTE
The works setting depends on the model of balance. The internal calibration modes are only available if
the self calibration system (SCS) is installed.
• SET WEIGHING