CIRAS-3 Operation Manual V. 1.09
170
I am unable to achieve my Tcuv set point
•
You are working outside the temperature control limits of 0-45
o
C.
•
The set point is not achievable because it is outside the temperature control range of ~10
o
C
below ambient to +15
o
C above ambient.
•
More than likely you are unable to cool the chamber to your set point as a result of high incident
light or your batteries are getting low.
I am unable to achieve my Tleaf set point
•
You are working outside the temperature control limits of 0-45
o
C.
•
The set point is not achievable because it is outside the temperature control range of ~10
o
C
below ambient to +15
o
C above ambient.
•
More than likely you are unable to cool the chamber to your set point as a result of high incident
light or your batteries are getting low.
PAR Measurement and Control
Why does my PARi reads lower than PARe
•
This is normal. The PARi sensor is located inside the cuvette head and below the window and
should therefore read approximately 10% lower than PARe.
•
Make sure that nothing is obstructing the PARi sensors.
I am using the light unit but I am unable to control light intensity
•
Check to make sure that the electrical connection is in place.
•
Go to Settings and make sure that you have the proper light source selected
My PARe readings do not seem accurate
•
The sensor likely requires recalibration. Check the calibration against another PAR sensor that
you trust and believe to be calibrated. Recalibrate if necessary. See
on page 106.