CIRAS-3 Operation Manual V. 1.09
165
Section 13. Troubleshooting & Diagnosis
System Power
The CIRAS-3 will not turn on when attempting to power up
•
The internal batteries are either missing or not charged. Connect the power supply adapter to the
CIRAS-3 and charge the batteries. Also check to make sure that the electrical connection is
intact.
•
Internal batteries are in “Ship” mode and need to be put into “Run” mode. See
on page 46.
The batteries are fully charged but the system can only operate for a couple of hours
and nowhere near normal expected times
•
You may be putting a high drain on the batteries (controlling temperature and/or light control)
which will have an impact on batteries. This is a fact of life unfortunately as temperature control
especially will put a large drain on the battery especially when cooling temperatures well below
ambient.
•
Old battery might be losing some juice over time and may require replacement.
I charged the batteries overnight but the batteries are not charged
•
Check the power supply adapter and when connected to mains power the indicating LED on it
should be a steady green. If it is not or if it is flashing it could mean a faulty charger. Also make
sure that your wall outlet or surge protector is on.
•
Make sure that the indicating LED on the console is showing a steady green indicating secure
connection.
•
Ensure batteries are seated properly in the battery compartment and the battery cable is
connected.
The EXT/PWR indicating LED on the CIRAS-3 console is flickering
It is possible that the CIRAS-3 was not powered off when the battery was extremely low and it continued
to drain beyond normal levels. If this happens the batteries may require a longer charging period to get
back to full charge status.