Step 1
. Remove CO
2
Cartridge holder and
regulator from main console by turning the holder
¼ counter clockwise and pull out.
Step 2
. Separate cartridge holder from regulator
as shown below.
Step 3.
Insert small coin in notch in regulator and
turn counterclockwise to remove inner fitting as
shown below.
Step 4
. Using small flat screwdriver, remove red
gasket seal and fit replacement making sure to
secure it in place around the piercing pin.
Re-Order Information
Part Number
Description
STD549
Cartridge Holder
STD546
Regulator Assembly
20577-1
Gasket CO
2
Seal
Cartridge Holder
Red Gasket
CO
2
Seal
Piercing Pin