All electrical and pneumatic connections are highlighted below. Remove the link pipe between “REF IN”
and “AIR OUT” which is only used when using the leaf cuvettes with the CIRAS-3.
1. Connect the 4 pin plug on the APA to the “Aux” socket on the CIRAS-3 console. Line the
arrow up on the black APA connector with the top middle of the “Aux” connector.
2. Connect the SRC-1 and CPY-4 to one of the 15 pin sockets on the APA and connect the
STP-1 Soil Temperature Probe to the other 15 pin socket on the APA (if used).
3. Connect the “Gas In” and “Gas Out” tubing on the SRC-1 and CPY-4 tubing to the “REF
IN” and “REF OUT” ports on the CIRAS-3 console.
Note. The CPY-4 is supplied with a water vapor equilibrator which we recommend is used for high
humidity conditions. The water vapor equilibrator should be plumbed as shown below: