![PP Systems CIRAS-3 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/pp-systems/ciras-3/ciras-3_operating-manual_1592668035.webp)
CIRAS-3 Operation Manual V. 1.09
35
PLC3 Leaf Cuvettes (Continued)
PAR Sensor (External)
Filtered, silicon cell (cosine corrected)
•
Response: 400-700 nm
•
Range – 0-3000 µmol m
-2
s
-1
•
Precision – 10 µmol m
-2
s
-1
Air Temperature Sensor
•
Range: -10
o
C to 50
o
C
•
Precision Thermistor
•
Accuracy - ± 0.5
o
C at 25
o
C
Temperature Control
Approximately 10
o
C below ambient to +15
o
C above ambient.
•
Temperature control limits – 0
o
C to 45
o
C
Leaf Temperature Sensor
PLC3 Universal
– Radiation sensor for non-contact measurement
•
Accuracy - ± 0.5
o
C at 25
o
C
PLC3 Narrow and Conifer
– Precision thermistor
•
Accuracy - ± 0.5
o
C at 25
o
C
Dimensions (General)
32 cm (L) x 3.8 cm (Handle diameter)
Weight
PLC3 Universal
– 0.750 kg
PLC3 Narrow and Conifer
– 1 kg
CFM-3 Chlorophyll Fluorescence Module
Modulating Beam
625 nm ± 5 nm (Red)
Saturation Light
0-10000 µmol m
-2
s
-1
Far Red Light
2 x 750 nm LEDs
Detector
PIN Photodiode with > 700 nm filter
Detector Method
Rapid pulse peak tracking
Leaf Area
1.75 cm
2
, 2.5 cm
2
and 4.5 cm
2
•
PP Systems is a registered trademark of PP Systems, Inc.
•
PP Systems is continuously updating its products and reserves the right to amend product
specifications without notice.
•
All brand names are trademarks of their respective owners.