
A-2
A TMS-A-305 Calibration Set
A.2 Defining the Depth Setting Standard
Before you can make a measurement using the depth setting standard, you
first of all have to define it, i.e. apply the data from the calibration certificate to
the file
Standards.xml
. If you want to use other standards, you also have to
define them first.
INFORMATION
If you define just one standard, you will not have to select a standard later
during calibration.
INFORMATION
To define depth setting standards, you require write permissions in the
directory
C:\Program Files\Polytec
.
To define the depth setting standard, you proceed as follows:
1. Open the file
Standards.xml
in a text editor. You will find the file in the
following directory:
C:\Program Files\Polytec\TMS
X.X\Pipeline\AddIns\CalibrationDataEvaluation