DÜRR CAS 1 suction system
Replace the coarse filter when it is full. Refer to the instructions supplied with
the DÜRR amalgam separator.
NOTE
You can grease the securing rings with vaseline to make them easier to
remove.
1. Remove the cover plate.
2. Remove the securing ring from the tube.
3. Remove the tube.
4. Remove the securing ring from the coarse filter.
5. Remove the coarse filter and empty it.
6. Repeat the procedure for the other coarse filter.
29 Cleaning and disinfection
244 Planmeca Compact i
User's manual