3. Select Service > Peripherals > ActiveAqua.
The
ActiveAqua maintenance window opens.
4. Press Reset next to
Replace prefilter in.
A confirmation message is displayed.
5. Confirm the reset by pressing OK.
6. Wait for 90 seconds while Planmeca ActiveAqua circulates water to
bleed the prefilter.
7. Press OK to close the window.
30 Planmeca ActiveAqua
User's manual
Planmeca Compact i 257