Steps
1. Open the
Apex locator window either by pressing the Apex locator button
or by activating the Morita TORX micromotor and selecting one of the
apex presets (a1 - a3).
2. Insert the tester into the jack on the back of the instrument console.
Compact i dental unit with hanging-tube instruments
Compact i dental unit with balanced instrument arms
3. Check that the meter indicates within 3 bars above or below 1.
The meter may jump when the tester is inserted. If it does, wait for about
one second until the meter stabilises and then check the reading.
If the reading is 4 or more bars away from 1, the unit will not make an
accurate measurement. Contact your Planmeca dealer.
23 Operating instruments
156 Planmeca Compact i
User's manual