1. Dentist’s area
2. Assistant’s area
35.3.2 Patient area
The patient area is 1.5 m (59.1”) in each direction from the dental unit.
The external PC, its keyboard and mouse, as well as Planmeca ProSensor’s
power supply and the Planmeca ProX generator assembly must be located
outside the patient area. The dentist, assistant and patient must not touch
the equipment outside the patient area during treatment.
CAUTION
Use only Planmeca specified devices inside the patient area.
CAUTION
The floor of the patient area must be dry.
NOTE
Connect only Planmeca specified devices to the dental unit.
NOTE
The external PC must be protectively earthed and IEC 60950 -approved (CE
marked).
NOTE
The monitor must fulfil IEC 60601-1 ed.3 requirements.
35 Technical information
306 Planmeca Compact i
User's manual