Separator system related error messages
ERROR
CODE
EXPLANATION
E63.1
Water management system (WMS) seems to be installed,
although it should not be (according to service mode).
E63.2
No Water management system (WMS) is installed, although it
should be according to service mode.
32.3.10 Patient chair related error messages
Patient chair related error messages
ERROR
CODE
EXPLANATION
E64
Lift motor position potentiometer or its cable is faulty or
disconnected.
E65
Lift motor does not run or no signal from position
potentiometer.
E65.1
Lift motor does not run. EmeStop relay PCB/cables or position
sensor/cable may be faulty.
E66
Lift motor or its position potentiometer cable has a wrong
connection.
E67
Backrest position potentiometer or its cable is faulty or
disconnected.
E68
Backrest motor does not run or no signal from position
potentiometer.
E68.1
Backrest motor does not run. EmeStop relay PCB/ cables or
position sensor/ cable may be faulty.
E69
Backrest motor or its position potentiometer cable has a
wrong connection.
E70.1
Lift motor error.
E70.2
Backrest motor error.
32.3.11 Main CPU related error messages
Main CPU related error messages
ERROR
CODE
EXPLANATION
E72.1
Program (Flash-EPROM) error. Main software is corrupted.
Replace Flash- EPROM.
E72.2
Program (Flash-EPROM) error. Boot software is corrupted.
Replace Flash- EPROM.
E72.3
Program (Flash-EPROM) error.
Update MCB software.
E72.4
Data transfer error while updating main PCB SW from control
panel memory device.
32 Help and error messages
286 Planmeca Compact i
User's manual
Содержание Compact i5
Страница 1: ...PlanmecaCompact i5 user s manual EN 30011782...
Страница 10: ...Table of contents Planmeca Compact i User s manual...
Страница 74: ...5 Replace the suction tubes in the reverse order 11 Suction system 64 Planmeca Compact i User s manual...
Страница 320: ...35 3 5 Side delivery with hanging tube instruments 35 Technical information 310 Planmeca Compact i User s manual...
Страница 325: ......