NOTE
You can also confirm the filling by pressing the Suction disinfectant button in
help message 47.1. See section "Filling suction disinfectant container" on
Steps
1. Press Program.
2. Press Maintenance.
The
Program - Maintenance window opens.
3. Press the Suction disinfectant button next to
Confirm container filling.
4. Press OK.
The pop-up window closes and help message H 47.2 is displayed to
inform that the suction disinfectant container is filled.
25.13 Wireless foot control
25.13.1 Pairing wireless foot control
About this task
If no wireless foot control is connected to your dental unit, or if you want to
change the current wireless foot control to another, you must pair the dental
unit and the foot control.
Steps
1. Press Program.
2. Press Foot control.
The
Wireless foot control window opens.
25 Programming
User's manual
Planmeca Compact i 201