information, see sections "Instrument spray" on page 181, "Automatic chip
blow" on page 182 and "Instrument light" on page 183.
CAUTION
Before modifying the torque and RPM functions, please refer to the file
manufacturer’s instructions concerning torque and RPM value
recommendations for the files.
Steps
1. Activate the micromotor.
2. Press Torque.
The micromotor's drive mode is automatically set to Auto stop. This
means, that when the torque limit is reached, the micromotor stops. No
other drive modes are available for the Bien-Air MCX micromotor.
3. Press Program.
4. Press RPM limit to adjust the micromotor speed rate.
A pop-up window opens.
5. Adjust the micromotor speed rate (RPM).
Alteration of RPM speed rate in different ranges
Range
In steps of:
100 - 200
10 (e.g. 110, 120 etc...)
200 - 500
20 (e.g. 220, 240, 260 etc...)
500 - 1 000
50 (e.g. 550, 600, 650 etc...)
1 000 - 2 000
100 (e.g. 1 100, 1 200 etc...)
2 000 - 5 000
200 (e.g. 2 200, 2 400 etc....)
5 000 - 10 000
500 (e.g. 5 500, 6 000, 6 500 etc...)
10 000 - 20 000
1 000 (e.g. 11 000, 12 000 etc....)
20 000 - 38 000
2 000 (e.g. 22 000, 24 000 etc....)
6. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
25 Programming
184 Planmeca Compact i
User's manual
Содержание Compact i5
Страница 1: ...PlanmecaCompact i5 user s manual EN 30011782...
Страница 10: ...Table of contents Planmeca Compact i User s manual...
Страница 74: ...5 Replace the suction tubes in the reverse order 11 Suction system 64 Planmeca Compact i User s manual...
Страница 320: ...35 3 5 Side delivery with hanging tube instruments 35 Technical information 310 Planmeca Compact i User s manual...
Страница 325: ......