161
enables the front-panel display if it has been turned off. This command will abort
a sweep or burst in progress.
-- S --
*SAV {1
︱
2
︱
3
︱
4}
Store the current instrument state to the specified non-volatile storage location.
•
If there is any instrument state stored in the specified location, it will be
overwritten.
•
You can store an instrument state into location 0 only from the remote
interface. It is not accessible from the front panel.
•
Storage location 0 is automatically overwritten when the power is cycled.
•
If you store the instrument state for an arbitrary waveform and then delete
it later, the waveform data is lost and the waveform generator is not able
to output the waveform.
•
When you store an instrument state, the selected function, frequency,
amplitude, dc offset, duty cycle, symmetry, as well as any modulation
parameters are saved also.
*TST?
Order a complete self-test of the waveform generator. Returns “0” when the test
runs successfully, or “1” when the test fails.
SWEep:SPACing {LINear
︱
LOGarithmic}
Specify the spacing for the sweep. The default sweep mode is Linear, which
waveform generator sweeps the frequency with spacing calculated in a linear
function. In logarithmic sweep mode, the waveform generator sweeps with
spacing calculated in a logarithmic function.
SWEep:SPACing?
Query the spacing used in sweeps. “LIN” or ”LOG” string indicating the spacing
will be returned.
SWEep:TIME {<seconds>
︱
MINimum
︱
MAXimum}
Specify the time duration in seconds to sweep from the start frequency to the
stop frequency. The range is from 1 ms to 500 seconds and the default is 1
second.