107
-222 Data out of range; PWM deviation limited by pulse
parameters
The PWM deviation (width or duty cycle) is adjusted to accommodate
the current pulse parameters such as pulse width/duty cycle, period
and edge time.
-222 Data out of range; duty cycle; value clipped to…
The square wave duty cycle is adjusted to within 20% to 80%.
-222 Data out of range; duty cycle limited by frequency; value
clipped to upper limit
The square wave duty cycle is adjusted to within 40% to 60% when
the frequency is greater than 10 MHz.
-223 Too much data
An arbitrary waveform specified contains more than 262,144 points.
Verify the number of points in the DATA VOLATILE or DATA:DAC
VOLATILE command.
-224 Illegal parameter value
An exact parameter value was expected.
7.3 Device Dependent Errors
-313 Calibration memory lost; memory corruption detected
A checksum error was detected in the non-volatile memory used to
store calibration constant.
-314 Save/recall memory lost; memory corruption detected
A checksum error was detected in the non-volatile memory used to
store instrument states.
-315 Configuration memory lost; memory corruption detected
A checksum error was detected in the non-volatile memory used to
store configuration settings.