5 Installation
Q-632 Miniature Rotation Stage
MP129E
Version: 1.1.2
23
Tools and accessories
If the Q-632 is to be operated in a vacuum:
−
E-873.UHVx extension cable, available as optional accessory (p. 10)
−
Suitable vacuum feedthrough, available as an optional accessory (p. 10):
C-815.VF (suitable for use in a vacuum to 10
-6
hPa)
−
Suitable tools for installing the vacuum feedthrough
Connecting the Q-632 to the electronics
1.
Prepare the Q-632 for connecting:
−
Remove the ESD protection from the Q-632's connector.
When a Q-632 is to be operated in a vacuum:
−
Installing the vacuum feedthrough:
a)
Obtain the dimensions from the corresponding dimensional drawing.
b)
Make sure that the vacuum feedthrough is aligned correctly:
C-815.VF: Vacuum side = D-sub 15 (f)
c)
Make a suitable opening in the vacuum chamber.
d)
Install the vacuum feedthrough.
2.
Connect the electronics to the Q-632 and if necessary, any further components as
shown in the connection diagram above.
3.
Take suitable measures to prevent the extension cable from being disconnected
accidentally.