SELF-DIAGNOSTIC FUNCTIONS
Alarms involving an error code indicated on the temperature display are generated by the unit’s
self-diagnostic function.
If any of the error codes are displayed, contact our sales representative or agent without delay, report the
error code, and request repairs.
Self-diagnostic
Cause
Indication
Buzzer
Remote alarm
Sensor
abnormality
The thermal sensor is disconnected.
Alarm indicator blinks.
E01 and -50 are displayed
alternately.
Intermittent tone
Alarm mode
The thermal sensor is
short-circuited.
Alarm indicator blinks.
E02 and 50 are displayed
alternately.
The defrost sensor is disconnected.
Alarm indicator blinks.
E03 and chamber temp. are
displayed alternately.
The defrost sensor is short-circuited.
Alarm indicator blinks.
E04 and chamber temp. are
displayed alternately.
The compressor sensor is
disconnected.
Alarm indicator blinks.
E05 and chamber temp. are
displayed alternately.
The compressor sensor is
short-circuited.
Alarm indicator blinks.
E06 and chamber temp. are
displayed alternately.
The ambient temperature sensor is
disconnected.
Alarm indicator blinks.
E07 and chamber temp. are
displayed alternately.
The ambient temperature sensor is
short-circuited.
Alarm indicator blinks.
E08 and chamber temp. are
displayed alternately.
The exclusive alarm sensor is
disconnected.
Alarm indicator blinks.
E13 and chamber temp. are
displayed alternately.
The exclusive alarm sensor is
short-circuited.
Alarm indicator blinks.
E14 and chamber temp. are
displayed alternately.
Battery switch
check (with
MPR-48B1)
When the battery switch is turned
ON without setting after attachment.
E09 and chamber temp. are
displayed alternately.
Intermittent tone
-----
Compressor
temperature
abnormality
The fan motor for cooling the
compressor is faulty.
Alarm indicator blinks.
E10 and chamber temp. are
displayed alternately.
Intermittent tone
Alarm mode
Battery check
(with MPR-48B1)
When about 3 years have passed
with the power supply cord
connected to the outlet.
(time to replace the battery)
F-1 and chamber temp. are
displayed alternately.
-----
-----
Fan motor check
When about 6 years have passed
with the power supply cord
connected to the outlet.
(time to replace the fan motor)
F-2 and chamber temp. are
displayed alternately.
-----
-----
The alarm mode of remote alarm terminal is cancelled by pressing the buzzer stop key (BUZZER) since
the remote alarm is operated in conjunction with buzzer. (except for power failure alarm status)