SETTING THE DELAY OF DOOR ALARM
The door check indicator is lit when the door is opened, and the buzzer sounds with some delay to notice
the door opening.
The delay time (between lighting of the door check indicator and activation of the buzzer) can be changed.
Set an appropriate delay time according to the condition of use to prevent the rise of chamber
temperature occurring as a result of the door being left open for too long.
Setting range of delay time: between 1 and 15 minutes
Initial setting (factory setting): 2 minutes
Example: Change the delay time to 3 minutes from 2 minutes
► Following shows a sample setting. Set the desired delay time according to the condition of use.
Operation
Key operated
Display after operation
1 -----
-----
The current chamber temperature is
displayed.
2
Press the up arrow key for 5
seconds.
F00 is displayed and the first digit
blinks.
3 Press the up arrow key 4 times.
The display is changed to F04 from
F00.
4 Press the “set” key.
SET
The current setting (002) is displayed
and the first digit blinks.
5
Press the up arrow key and scroll
the figure to 3.
The display is changed to 003 from
002.
6 Press the “set” key.
SET
Delay time is memorized and the
current chamber temperature is
displayed.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.