21
SETTING THE TEMPERATURE DISPLAY MODE
The display mode of chamber temperature is selectable; integer display or decimal display. Select the
display mode as required.
Setting range of temperature display mode: integer mode (display: -50 to 50) or
decimal display (display:-19.9 to 50.0)
Initial setting (factory setting): integer display mode
In the case of decimal display mode, -19.9 is always displayed when the chamber temperature is lower
than -19.9
o
C.
Example: Change the mode to decimal display from integer display (factory setting)
Operation
Key operated
Display after operation
1
-----
-----
The current chamber temperature is
displayed.
2
Press the up arrow key for 5
seconds.
F00 is displayed and the first digit
blinks.
3
Press the up arrow key 8 times.
The display is changed to F08 from
F00.
4
Press the scroll key once.
The second digit blinks.
5
Press the up arrow key once.
The display is changed to F18 from
F08.
6
Press the “set” key.
SET
The current setting (000) is displayed
and the first digit blinks.
7
Press the up arrow key and scroll
the figure to 1.
The display is changed to 001 from
000.
8
Press the “set” key.
SET
The setting is memorized and the
current chamber temperature is
displayed with decimal mode.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.
Note:
Set the figure to 000 in above procedure 7 when selecting the integer display mode.