23
SETTING THE ALARM TEMPERATURE
Setting the high-temperature alarm
An abnormality (chamber temperature rise) is indicated by the blinking of red LED lamp and the chamber
temperature display and the buzzer sounding (15 minutes after blinking) if the chamber temperature
exceeds the set value of the high-temperature alarm. Always set the high-temperature alarm to protect
stored items from damage resulting from an increase in temperature.
Setting range for high-temperature alarm:
between 2
o
C and 14
o
C higher than the chamber set temperature
Initial setting (factory setting): 5
o
C higher than the chamber set temperature
Example: Change the high temperature alarm to chamber temperature plus 3
o
C from chamber
temperature plus 5
o
C
► Following shows a sample setting. Set the desired high-temperature alarm according to the condition
of use.
Operation
Key operated
Display after operation
1 -----
-----
The current chamber temperature is
displayed.
2
Press the up arrow key for 5
seconds.
F00 is displayed and the first digit
blinks.
3 Press the up arrow key once.
The display is changed to F01 from
F00.
4 Press the “set” key.
SET
The current setting (005) is displayed
and the first digit blinks.
5
Press the up arrow key and scroll
the figure to 3.
The display is changed to 003 from
005.
6 Press the “set” key.
SET
Set temperature is memorized and
the current chamber temperature is
displayed.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.
<Important>
The high temperature alarm may be activated after defrosting or a large amount of load is stored in the
chamber. This is not a malfunction. The high temperature alarm is cancelled automatically when the
chamber temperature reaches the set temperature.