18
PLACEMENT OF STORED ITEMS
Do not place stored items so as to disrupt the air circulation in the chamber. Disruption of air circulation
can cause the freezing of stored items or can impair the temperature distribution in the chamber.
Air intake vent
Do not block the air intake vent with the stored items. Care should be taken not to put any paper or plastic
chips near the air intake vent. These may be sucked in and block the air intake vent.
Air exhaust vent
Do not block the air exhaust by the stored items near the air exhaust vent. Care should be taken not to
place stored items near to the air exhaust vent. The stored items may become frozen when the
refrigerator is operating at temperature settings lower than 5
o
C.
Load line [see figure below]
The storage area is limited to ensure adequate air flow in the chamber. The label “load line” is attached to
the right and left sides of the chamber. Always place stored items within the area limited by these labels.
Load line
Load line
Stock space
Side view
Cooling air
Fan
Air exhaust vent
Air intake vent