12
INSTALLATION
1.
After unpackaging
Remove all transportation packaging materials and tapes. Open the doors (refrigerator and freezer) and
ventilate the unit. If the outside panels are dirty, clean them with a diluted neutral dishwashing
detergent.
Undiluted detergent can damage the plastic components. For the dilution, refer to the instruction of
the detergent.
After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the
panels with a dry cloth.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
2.
Installing the unit
Extend the leveling feet by rotating them counterclockwise
until 2 front casters separate from the floor by 5 mm to 10
mm. [Fig. 1]
And adjust the leveling foot to level the unit.
The unit is installed firmly by separating the front casters
from the floor. The unit may be moved at the time of
opening/closing the door if the front casters contact with
floor.
3. Fixing the unit
Fix the unit to the wall by using 2 back spacers and a rope
or chain. [Fig. 2]
4. Ground (earth) the unit
The ground (earth) is for preventing the electric shock in
the case of the electrical insulation is somehow degraded.
Always ground the unit at the time of installation.
For a 3-prong plug with grounding pole, there is no need
for electric work for grounding.
If the power supply outlet is not grounded, it will be
necessary to install a ground by qualified engineers.
5. Connection of remote alarm equipment
The alarm status is noticed to a remote location when a
remote alarm equipment (commercial item) is connected to
the remote alarm terminal. It is recommended to install a
remote alarm equipment (commercial item) when the
refrigerator is installed in a desolate location so that an
alarm status is noticed to an operator.
Contact our sales representative or agent for the
installation of a remote alarm equipment (commercial
item).
Leveling foot
Fig. 1
Back spacer
Fig. 2