9
REFRIGERATOR COMPONENTS
1. Control panel:
With the key on the control panel, temperature display can be changed and the alarm
buzzer can be canceled. Also, the running status can be checked on the temperature display and the
indicators. [page 10]
2. Keyhole:
The door of refrigerator and freezer can be locked with the enclosed key.
3. Drain port (bottom of freezer):
Remove the cap when the water resulting from defrosting is drained.
The drained water is evaporated in the evaporating tray under the freezer.
Replace the cap after drainage.
4. Leveling foot (front, 2 locations):
These are screw bolts used to install and fix the unit. Adjust the
height of the leveling feet by turning the screw bolts until 2 front casters are away from the floor.
[page 12]
5. Space for a temperature recorder:
An optional temperature recorder is installed here [page 34]. For
the installation of temperature recorder, contact our sales representative or agent.
6.
Shelf (for freezer):
Use to store the materials in the freezer. It is recommended to store the items on
the shelf. The maximum load is 10 kg.
Never touch the storage items with wet hands. Touching with the wet hands may cause frostbite.
7. Access port (left side):
This port allows a sensor or cable of measuring equipment to enter the
chamber from outside. Also, the capillary tube of temperature recorder is passed through the port.
Replace the insulations and the rubber caps when the access port is not used. Improper replacement
may cause rise of chamber temperature or condensation around the access port.
8.
Air intake vent (lower back):
An intake vent for circulating the air in the refrigerator chamber. Do not
block this vent. Blocking this vent may cause unstable chamber temperature. Do not insert fingers or
similar articles into the vent.
9.
Shelf (for refrigerator):
Use to store the items in the refrigerator. It is recommended to store the
items on the shelf. The maximum load is 20 kg per shelf.
10.
Air exhaust vent (center back):
An exhaust vent for circulating the air in the refrigerator chamber.
Do not block this vent. Blocking this vent may cause unstable chamber temperature. Arrange the
storage items not to subject to the cold air from this vent to avoid freezing.
11. Power switch (also served as a circuit breaker):
Remove 2 screws fixing the cover and the power
switch that is also served as a circuit breaker is accessible. The round button on the left of the power
switch is a check button for circuit breaker. The operation of the circuit breaker is checked by pressing
this button.
Please be advised that the power to the unit is disconnected when the check button is pressed.