13
START-UP OF UNIT
Follow the procedures below when starting-up (initial and after stop due to the cleaning, maintenance or
moving) of the unit.
At the recovery from power failure, the operation is start-up automatically with the setting before power
failure.
1.
With the refrigerator and freezer empty, connect the power supply cord to the dedicated outlet having
an appropriate rating, and turn on the power switch.
► The blinking refrigerator temperature is displayed on the temperature display and the alarm indicator
blinks. (This is caused by high temperature alarm function and not a malfunction.)
<Important>
5 minutes is needed to re-start the compressor. Keep 5 minutes before turning on the power switch
when the power switch is turned off.
2.
The refrigerator temperature is set to 5
o
C and the freezer temperature is set to -30
o
C at the factory.
Set the temperature of refrigerator or freezer when other temperature setting is desired. [page 14]
The storage items may be frozen practically when the refrigerator temperature is set to 3
o
C or lower.
3.
On the temperature display check the refrigerator temperature reaches 5
o
C and the freezer
temperature reaches -30
o
C, or reaches the desired temperature
4.
Check the lamp in the refrigerator is on by opening the refrigerator door.
5.
Begin slowly placing items into the refrigerator and freezer to minimize the temperature rise.
<Important>
Do not block the air intake vent or air exhaust vent in the refrigerator. Arrange the storage items with
adequate space to keep the air circulation. Too much storage items in the refrigerator makes the
temperature around the air exhaust vent at -2
o
C when the refrigerator temperature is set to 2
o
C. It is
recommended to set the refrigerator temperature to 4
o
C or 5
o
C when the items that should not be frozen
are stored.
<Attention>
The refrigerator door has a condensation on the glass during operation. Wipe off the condensation with
a dry soft cloth.