DECONTAMINATION
When the chamber is contaminated or when cleaning the chamber prior to starting a culture, it is possible
to perform H
2
O
2
decontamination.
H
2
O
2
decontamination function is workable when H
2
O
2
generator MCO-HP, H
2
O
2
decon board
MCO-170HB and electric lock MCO-170EL are installed. If not, H
2
O
2
decontamination can not be
performed.
When an optional small door (MCO-170ID) is installed, H
2
O
2
decontamination can be performed with an
MCO-170ID.
WARNING
Use the reagent specified by our company for H
2
O
2
decontamination
. Using a different H
2
O
2
solution
may cause explosion or damage to the incubator, or insufficient decontamination.
Do not use chemicals other than the H
2
O
2
reagent, such as alcohol.
Doing so may result in damage to
the H
2
O
2
vapor generator.
WARNING
When performing H
2
O
2
decontamination,
make sure that the outer and inner doors are securely
closed
. During H
2
O
2
decontamination, plug the access hole with the silicon caps that are provided. Failure
to do so may be harmful to health due to leakage of H
2
O
2
gas.
CAUTION
H
2
O
2
decontamination can be performed only for the chamber and inner attachments with standard
specifications, and not for any other objects.
CAUTION
Perform H
2
O
2
decontamination with the inner attachments arranged as specified by our company
.
Arranging them in a different way may result in insufficient decontamination.
CAUTION
Wear rubber gloves when handling the H
2
O
2
reagent
. Direct contact with the H
2
O
2
reagent may result in
inflammation of the skin.
CAUTION
After H
2
O
2
decontamination has been completed, residual H
2
O
2
solution will remain on the bottom of the
chamber, the H
2
O
2
vapor generator, and the bottom of the duct.
Wearing protective glasses and rubber
gloves, wipe it off with a non-woven cloth
. Failure to do so may result in a deficient culture.