29
CORRECT OPERATION
Use the following procedure to start trial operation or actual operation of the incubator.
1.
Install the incubator correctly, referring to “INSTALLATION” on page 16 to 21.
2.
Remove the packing materials from the chamber and inner attachments. Clean and disinfect the
chamber and all the inner attachments, referring to “ROUTINE MAINTENANCE” on page 82.
3.
Add approximately 1.5 L of sterile distilled water to the humidifying pan (Refer to page 26).
4.
Connect the removable power supply cord that is provided, to the port on the lower rear side.
5.
Connect the removable power supply cord to the outlet.
6.
Turn ON the power switch on the lower right side of the incubator.
7.
Set the frequency of a power supply on the LCD touch panel (Refer to page 58~59).
WARNING
Always use the removable power supply cord that is provided.
Other power supply cord may cause
electric shock or fire.
The provided removable power supply cord is only for this product.
Never use it for any other products.
When the incubator is not in use
Empty the water out of the humidifying pan and remove moisture from the chamber. Make sure that the
chamber is completely dry before closing the doors. Failure to do so may result in damage.
Before moving the incubator
Before moving the incubator, empty the water out of the humidifying pan, disconnect the power supply plug
from the outlet, and make sure that the cord will not be damaged. Failure to do so may result in electric
shock or fire.