decontamination
1.
Take out all the trays, the fan cover, the duct, the humidifying pan cover and the humidifying pan from
the chamber.
2.
Dispose of the water in the humidifying pan, and wipe inner attachments removed from the chamber,
with a gauze containing water or alcohol for disinfection.
3.
Wipe the inside walls of the chamber with a gauze containing water or alcohol for disinfection.
4.
Attach the duct and the fan cover.
5.
Pour one bottle of the H
2
O
2
reagent MCO-H2O2
into the H
2
O
2
generator MCO-HP (Fig. 1).
6.
Set the two pins on the H
2
O
2
generator in the 2
holes on the lower left side of the duct (Fig. 1).
Note:
Make sure that the H
2
O
2
generator is
securely attached. Uncertain attachment may result
in insufficient decontamination.
7.
Remove a connector cap from connector on
bottom right of the far side of the chamber. Connect
the H
2
O
2
generator and connector by a cable that is
provided (Fig. 2).
Note:
Be sure to keep the connector cap.
8.
Insert 4 trays in the 2nd, the 3rd, the 4th and the 6th tray catches from the top of the chamber side.
Note:
The trays included as accessory are designed to be appropriate for decontamination. If half tray
(MCO-25ST, optional) or trays for previous models are used, decontamination may not be sufficiently
effective.
H
2
O
2
generator
Duct lower part
Fig. 1
Pour one bottle of
the H
2
O
2
reagent
Fig. 2
Connector
Front