19
4. Adjust the leveling feet.
Adjust the leveling feet by turning them
counterclockwise to level the incubator (Fig. 3).
Note:
Incubating on a leaning tray may have a bad
influence on the cultivation.
5. Ground the incubator.
Ground the incubator during installation to prevent electric shock in case the insulation is not sufficient. If
there is no ground wire at the location, consult with qualified service personnel.
When a ground must be installed
If a grounded 3-pole outlet is not available, then a ground must be installed. Consult with qualified service
personnel.
WARNING
To prevent electric shock,
always ground the incubator.
If grounding is not possible, then have a ground
installed by qualified personnel. If the incubator is not grounded, it may result in electric shock.
Never connect the ground wire to a gas pipe, water pipe, lightning rod, or telephone ground wire.
Doing so may cause electric shock.
Installing a ground fault circuit breaker
If using the incubator in the location with moisture or humidity cannot be avoided, then it is recommended
that a ground fault circuit breaker be installed in the power supply circuit (i.e., the power supply at the
incubator). Have the circuit breaker installed by qualified service personnel.
CAUTION
Do not climb on the incubator or place objects on top of it.
Doing so may damage it or cause it to fall
over, resulting in injury.