ALARM PARAMETERS
Automatic set temperature alarm delay (Temp. Alarm Delay):
The function is that when the incubator is in the alarm state of automatic set temperature, the alarm buzzer
will sound after the automatic set temperature alarm delay set time passed.
Settable range: 0 minute
~
15 minutes, factory setting: 15 minutes.
Note:
When the incubator is recovered from the alarm state of automatic set temperature within the
automatic set temperature alarm delay time, the buzzer doesn’t sound after the elapse of the automatic set
temperature alarm delay.
·
CO
2
Alarm:
When the chamber CO
2
density exceeds the scope, the set CO
2
density ± the set value of Automatic set
CO
2
density alarm, the alarm is activated. Settable range: ±0.5 %
~
±5.0 %, factory setting: ±1.0 %.
·
Automatic set CO
2
density alarm delay (CO
2
Alarm Delay):
The function is that when the incubator is in the alarm state of automatic set CO
2
density, the alarm buzzer
will sound after the automatic set CO
2
density alarm delay set time passed.
Settable range: 0 minute
~
15 minutes, factory setting: 15 minutes.
Note:
When the incubator is recovered from the alarm state of automatic set CO
2
density within the
automatic set CO
2
density alarm delay time, the buzzer doesn’t sound after the elapse of the automatic set
CO
2
density alarm delay.
5.
On the Alarm Setting #2 screen, it is possible to
set each alarm. Press Apply key to save the input
value and setup. The display returns to the Tools #1
screen.
●
Each setting
·
Door Delay:
The function is that when the incubator is in the alarm state of door, the alarm buzzer will sound after the
alarm delay set time passed. Settable range: 1 minute
~
30 minutes, factory setting: 2 minutes.
Note:
When the incubator is recovered from the alarm state within the door alarm delay time, the buzzer
doesn’t sound after the elapse of the door alarm delay.
·
Ring Back:
The function is that the alarm buzzer sounds again when the alarm state still continues after the alarm
delay set time passed even though the alarm buzzer was stopped by pressing Buzzer key. By holding and
sliding Ring Back slide key to the right, the Ring Back is turned to ON. Settable range: 1 minute
~
99
minutes, factory setting: 30 minutes.
Note:
At Err01 (CO
2
gas cylinder empty), Err11•12 (CO
2
sensor error), Err18 (UV lamp failure) and Door
alarm, the alarm is not re-activated because the alarm itself is deactivated by pressing Buzzer key (refer to
page 83 and 84).
·
Remote Alarm:
The function is that the remote alarm is continued even though the buzzer is stopped by pressing Buzzer
key. By holding and sliding Remote Alarm slide key to the right, the Ring Back is turned to ON (not in
conjunction with Buzzer key). Factory setting: ON.
6.
(From procedure
4
and
5
) Press Top key to return to the Top screen.