64
H
2
O
2
DECONTAMINATION
9.
Set the humidifying pan cover, the humidifying pan removed in the procedure
1
(Fig. 3).
Note:
H
2
O
2
decontamination can be performed only for the chamber and inner attachments with standard
specifications, not for any other objects.
10.
Make sure that the duct, the fan and the H
2
O
2
generator are securely attached (refer to Fig. 1 to 3 on
page 88). After that, close the inner door and the outer door.
Note:
When an optional small door (MCO-170ID) is installed, make sure that 4 gastight split doors are
closed.
11.
Press H
2
O
2
key for 3 seconds to open Setting
Position window.
Note:
·
When the model does not have H
2
O
2
decontamination function, H
2
O
2
key is not
displayed on the Top screen.
·
When key lock is ON, Password input window is
opened and input of the release password of Key
Lock is required. Refer to page 37.
12.
Finally make sure that the inner attachments
and the H
2
O
2
generator are attached correctly. After
confirm, press Next key to lead the H
2
O
2
Decontamination screen.
Fig. 3
Front
Humidifying
pan cover
Humidifying pan
H
2
O
2
generator