43 

 

 

4.

 On the Setting screen, input Log Interval. Press 

Apply key to save the input value. The display 
returns to the Log screen.

 

Settable range: 2 minutes

~

30 minutes. 

Factory setting: 6 minutes.

 

 
Note:

 It is possible to register 8-digit alphanumeric 

characters as the Unique ID. Refer to page 48. 

 
 
 
Note: 

Relation between log interval and the estimated amount of data that can be saved 

Log interval=2 minutes: Approx. 46 days 
Log interval=6 minutes: Approx. 135 days 
Log interval=30 minutes: Approx. 664 days 

When saving data more than the above, and the data is overwritten and the old data is delated. 

 
5.

 Press Top key to return to the Top screen. 

 

  Displaying operation log 

 

Operation log saved in the incubator can be displayed graphically on the LCD touch panel. 
 

1.

 Press Menu key to lead the Menu screen.

 

 
 
 
 
 
 
 
 
 
 
2.

 Press Log key to lead the Log screen. 

 
 
 
 
 
 
 
 
 
 

Содержание MCO-170AICUVL

Страница 1: ...carefully before using this product and keep the operating instructions for future use See page 92 for model number Operating Instructions CO2 Incubator This product is for research use only Not for...

Страница 2: ......

Страница 3: ...nts 25 Filling humidifying pan 26 FOR BETTER CULTIVATION Precautions for cultures 27 Preventing contamination 28 CORRECT OPERATION 29 BASIC OPERATION ON LCD TOUCH PANEL 30 BASIC PARAMETERS Numerical i...

Страница 4: ...LOCK OPTION 68 Setting User ID 68 Setting auto lock 70 Using unlock key 73 Removing auto lock 74 GAS AUTO CHANGER OPTION 75 Connecting CO2 gas cylinder 75 Automatic CO2 gas supply line changeover 76 M...

Страница 5: ...Contact our sales representative or agent if any page of the operating instructions is lost or the page order is incorrect Contact our sales representative or agent if any point in the operating inst...

Страница 6: ...lure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Symbol shows This symbol means caution This symbol means an action is prohibited This s...

Страница 7: ...and safe disposal options visit ec gc ca mercure mercury Dispose of or recycle in accordance with applicable laws Pour plus de renseignements sur les proc dures de manutention s curitaire les mesures...

Страница 8: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 9: ...cord Pulling the cord may result in electric shock or fire by short circuit Disconnect the power supply plug before moving the unit Take care not to damage the power supply cord A damaged cord may ca...

Страница 10: ...of operation after power failure or turning off of power switch The stored items may be damaged due to the change of setting Be careful not to tip over the unit during movement to prevent damage or in...

Страница 11: ...prevent electric shock Only a qualified engineer or service personnel should be allowed to open these covers This symbol indicates an ultraviolet light UV caution This symbol indicates that caution i...

Страница 12: ...C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvolta...

Страница 13: ...an optional electric lock MCO 170EL is installed When 13 is removed Sample air outlet cap 18 17 16 15 14 Lower right side Removable power supply cord port 20 21 Access port cover Silicon cap Rear sid...

Страница 14: ...nd humidity recovery 10 Humidifying pan Fill the humidifying pan with sterile distilled water and set the humidifying pan with the inner side flush against the back Install the humidifying pan in a lo...

Страница 15: ...2 density is set 0 5 Set CO2 density value display field The set value of the chamber CO2 density is displayed Default setting 0 6 CO2 gas injection indicator This lamp lights when CO2 gas is being in...

Страница 16: ...ly in normal characters and reverse video Close Door Closed is displayed Locked Door Locked is displayed 2 13 Message display field Alarms errors or messages are displayed when fault occurs Refer to p...

Страница 17: ...be informed at a remote location from this product by connecting the external alarm unit to the remote alarm terminals For the type and behavior of remote alarm output refer to page 83 to 85 The termi...

Страница 18: ...lers ovens or autoclaves Heat will adversely affect the performance of the incubator Ambient temperature at least 5 C lower than set temperature The control temperature of the incubator is at least 5...

Страница 19: ...In addition do not install it near water or steam pipes Moisture can cause the insulation to deteriorate which may result in current leakage or electric shock No inflammable or corrosive gas Never in...

Страница 20: ...e tie banding the power supply cord Prolonged banding may cause the corrosion of the cord coating WARNING Do not leave the plastic wrapping bags within reach of children If the bag is placed over a ch...

Страница 21: ...ys ground the incubator If grounding is not possible then have a ground installed by qualified personnel If the incubator is not grounded it may result in electric shock Never connect the ground wire...

Страница 22: ...O2 CO2 incubator other than this product use the stacking plate MCO 170SB Refer to table 11 on page 91 When the incubator is not in use Empty the water from the humidifying pan and remove moisture fro...

Страница 23: ...tube tightly to prevent it from coming off Make sure that the gas tube is not folded If the CO2 gas is supplied to multiple CO2 incubators from a single gas cylinder a CO2 solid will be formed in the...

Страница 24: ...n corrosion or rusting Be careful to keep the diluted detergent or water out of the temperature sensor the CO2 gas injection port the inner sample air access port the fan motor shaft bearing and the i...

Страница 25: ...arp edges or corners Be careful not to damage the UV lamp in the duct 1 Turn OFF the power to the incubator 2 Open the outer and inner doors and pull out all the trays Fig 1 3 Pull out the humidifying...

Страница 26: ...6 Lift the duct and remove it from the pins on the rear side Fig 5 7 Pull out the fan Fig 6 8 Remove the silicon caps of each access port from interior Fig 7 and exterior Fig 8 Duct Fig 5 Fig 6 Fig 7...

Страница 27: ...sition the long hole of duct with the projection of fan cover and insert directly Fig 3 Same applies for the humidifying pan cover The fan cover may lean by strongly pushing the head of it in the back...

Страница 28: ...l direction as its shorter side is placed in the back Refill the humidifying pan with water early when the volume of water is decreased Mixing any reagent in the water for humidifying may have a bad i...

Страница 29: ...re medium incomplete closing or damage to the gasket Before opening the inner door check through the glass to confirm that the UV lamp is OFF Be careful when closing the outer door Use the handle when...

Страница 30: ...n taking them in and out Keep the chamber clean Wipe off any fingerprints If water spills from the humidifying pan or if the doors are left open for a long time condensation may form on the inside of...

Страница 31: ...switch on the lower right side of the incubator 7 Set the frequency of a power supply on the LCD touch panel Refer to page 58 59 WARNING Always use the removable power supply cord that is provided Oth...

Страница 32: ...rm Export Export Alarm log 51 53 Lock Lock screen Auto Lock Auto Lock screen 1 Setting Auto lock ON OFF User ID ON OFF 70 74 KeyLock Key Lock screen Setting Key lock ON OFF password 35 38 Auto Lock Us...

Страница 33: ...n are changed inaccurate temperature or CO2 density may be displayed Operation from other than Menu key CO2 gas supply line select key Operate Manual changeover between CO2 gas supply line A and B Pag...

Страница 34: ...e and press OK key Key description Numerical key 0 9 Input numerical values Up Down key Increases or decreases the numerical value displayed in the numerical input box Clear key Deletes the numerical...

Страница 35: ...CO2 density and the temperature of the high limit temperature alarm for normal operation according to the following procedure The incubator automatically starts operation using these settings after po...

Страница 36: ...f each alarm Note When changing the set temperature from less than 45 0 o C to 45 0 o C or higher the incubator readjusts the CO2 sensor During readjusting Status Gas sensor initializing is displayed...

Страница 37: ...35 Setting key lock 1 Press Menu key to lead the Menu screen 2 Press Lock key to lead the Lock screen 3 Press KeyLock key to lead the Key Lock screen...

Страница 38: ...ight Key Lock turns to ON Password The number Max 6 digit inputted here are registered the release password of Key Lock Confirm Password To prevent erroneous input input the same password as Password...

Страница 39: ...t of the release password of Key Lock is required When the inputted password is incorrect Notice dialog box is displayed Press OK key and then input the correct password Removing key lock 1 By pressin...

Страница 40: ...Lock screen 5 On the Key Lock screen by holding Key Lock slide key and sliding to the left change to OFF Press Apply key to turn the key lock OFF The display returns to the Lock screen Note The releas...

Страница 41: ...ocedure 4 Press Alarm Setting 2 key to lead the Alarm Setting 2 screen it is possible to set door alarm delay ring back and remote alarm go to procedure 5 4 On the Alarm Setting 1 screen input each pa...

Страница 42: ...nsity alarm delay 5 On the Alarm Setting 2 screen it is possible to set each alarm Press Apply key to save the input value and setup The display returns to the Tools 1 screen Each setting Door Delay T...

Страница 43: ...e behavior at the time of pressing Buzzer key and the re activation of alarm under each setting condition refer to Table 5 7 on page 84 Resolve the cause of the alarm in reference to page 83 85 becaus...

Страница 44: ...ation log data chamber temperature CO2 density and open close state of outer door Use the following procedure to set the log interval interval of acquiring the operation log 1 Press Menu key to lead t...

Страница 45: ...interval and the estimated amount of data that can be saved Log interval 2 minutes Approx 46 days Log interval 6 minutes Approx 135 days Log interval 30 minutes Approx 664 days When saving data more...

Страница 46: ...isplay graphically 5 On the Chart screen by pressing Show key after pressing the item you want to display graphically the graph of each operation log is displayed Actual Temp Chamber temperature log g...

Страница 47: ...turn to the Top screen 8 Door Opening log graph is displayed Press Back key to return to the Chart screen Press Top key to return to the Top screen Note The error of about 1 minute may be observed dur...

Страница 48: ...to the USB memory inserted into the USB port 1 Insert the USB memory into the USB port Note It is not possible to use a USB memory with security functions that requires entering password 2 Press Menu...

Страница 49: ...type of operation log data you want to export To export all types of operation log data press All Ch key To export only operation log data you want to export select operation log data you want to expo...

Страница 50: ...s created in the USB memory and the exported file is saved in it in CSV format The exported file name is in date 8 digits type of data format e g When exporting all types of data using All from Oct 1s...

Страница 51: ...Alarm log saved in the incubator can be displayed graphically on the LCD touch panel Note When saving alarm logs more than 257 the oldest alarm log is deleted and then overwritten 1 Press Menu key to...

Страница 52: ...e 1 day 45 days Note The error of about 1 minute may be observed during 1 month Refer to page 54 for the procedure of setting time Press Back key to return to the Log screen Press Top key to return to...

Страница 53: ...the USB port by CSV format 1 Insert a USB memory in the USB port Note It is not possible to use a USB memory with security functions that requires entering password 2 Press Menu key to lead the Menu s...

Страница 54: ...button and input days Settable range 1 day 45 days Note The error of about 1 minute may be observed during 1 month Refer to page 54 for the procedure of setting time 6 Press Export key Note When USB m...

Страница 55: ...deleted 8 Remove a USB memory from the USB port Note A log folder is created in a USB memory and an exported data file is saved in the log folder by CSV format Exported file name The first date durin...

Страница 56: ...key to lead the Date Time screen 4 On the Date Time screen input the present date and time Press Apply key to save the input value The display returns to the Tools 2 screen Note 24 hour clock It is re...

Страница 57: ...Tools 2 key to lead the Tools 2 screen 3 Press Brightness Sleep key to lead the Brightness Sleep screen 4 On the Brightness Sleep screen each setting of brightness and sleep is available Press Apply...

Страница 58: ...right the Sleep function is turned to ON Input the set value of time to change the Sleep state Settable range 1 minute 5 minutes factory setting 2 minutes Note It is not possible to operate any key in...

Страница 59: ...stribution because the UV lamp generates heat for a long time It may also shorten the service life of the UV lamp The preset period can be changed when necessary as shown in the page 58 59 To check wh...

Страница 60: ...information on replacing the UV lamp If the UV lamp burns out Err18 UV Lamp Abnormal is displayed in the message display field it is not possible to perform H2O2 decontamination Replace the UV lamp a...

Страница 61: ...lamp has turned on is displayed as the percentage to 5 000 hours which are recommendation time to replace It is impossible to set UV Timer Ext The more total time which UV lamp has turned on increases...

Страница 62: ...the chamber including the trays the fan cover the duct the fan the humidifying pan and the humidifying pan cover Clean all the attachments in an autoclave or with alcohol for disinfection 2 Clean and...

Страница 63: ...n display when UV lamp is lighting Note The UV 24 hour mode may cause the automatic set temperature alarm because of a rising chamber temperature After procedure 8 by opening the outer door when UV la...

Страница 64: ...contamination make sure that the outer and inner doors are securely closed During H2O2 decontamination plug the access hole with the silicon caps that are provided Failure to do so may be harmful to h...

Страница 65: ...n the lower left side of the duct Fig 1 Note Make sure that the H2O2 generator is securely attached Uncertain attachment may result in insufficient decontamination 7 Remove a connector cap from connec...

Страница 66: ...an optional small door MCO 170ID is installed make sure that 4 gastight split doors are closed 11 Press H2O2 key for 3 seconds to open Setting Position window Note When the model does not have H2O2 de...

Страница 67: ...zer sounds when H2O2 decontamination is completed About 100 minutes later CAUTION The electric lock will remain locked if power supply is cut off during H2O2 decontamination After the power supply is...

Страница 68: ...and inner doors Disconnect connector of the chamber cable and H2O2 generator and remove H2O2 generator and cable from the chamber Note When doing the above work put on protection glasses and rubber g...

Страница 69: ...er cover securely to prevent impurities from becoming mixed in the reagent Check the container to make sure that there is no damage corrosion or cracking Store the container with the inlet facing upwa...

Страница 70: ...king the outer door are as follows Quick mode Press the Unlock key User ID mode Input the User ID and release password of Auto Lock after pressing the Unlock key Setting User ID Before turning the Use...

Страница 71: ...ord as Password input box When inputting different password Notice dialog box is displayed Press OK key and input the correct password Note A release password of Auto Lock is for unlocking the outer d...

Страница 72: ...registered User ID into User ID input box and input its registered password into Password input box Press Delete key to delete the registered User ID and its password Note When deleting all registered...

Страница 73: ...of unlocking the outer door between the quick mode or the User ID mode By holding the User ID slide key and sliding it right the User ID mode is turned to ON Factory setting OFF quick mode Note When n...

Страница 74: ...ID and its release password of Auto Lock Note The User ID which is inputted at this time is saved as the open close state of outer door log data refer to page 46 47 When the inputted User ID or its p...

Страница 75: ...ed use the unlock key that is provided To re lock the outer door turn the unlock key to the lock direction while the outer door is open Close the outer door after the out door is locked manually Note...

Страница 76: ...d the Lock screen 3 Press Auto Lock key to lead the Auto Lock screen 4 On the Auto Lock screen by holding the Auto lock slide key and sliding it left the Auto lock is turned to OFF Press Apply key to...

Страница 77: ...he CO2 gas cylinder A 3 Using a gas tube that is provided connect the connecting port B for CO2 gas pipe and the gas regulator of the CO2 gas cylinder B Note If the CO2 gas is supplied to multiple CO2...

Страница 78: ...o changer MCO 21GC is installed CO2 gas supply line indicator A B and CO2 gas supply line select key are displayed in the Top screen CO2 gas supply line indicator A or B being used is lighted CO2 gas...

Страница 79: ...as cylinder A Table 2 CO2 gas supply line automatic changeover e g When CO2 gas cylinder A is empty it changes over CO2 gas cylinder B Situation CO2 gas CO2 gas supply line indicator Message display f...

Страница 80: ...supply line A is changed to B by CO2 gas automatic changer function the CO2 gas supply line B is changed to A manually without pressing the Buzzer key Table 3 Situation CO2 gas CO2 gas supply line ind...

Страница 81: ...2 incubator Since a Standard gas cylinder is used as a standard of exact density during CO2 density calibration prepare a standard gas cylinder that is same as the set CO2 density Note There is not th...

Страница 82: ...ettable range 4 0 21 0 Factory setting 5 0 Note Don t turn OFF CO2 STD Gas Density slide key 6 Press STD Gas Calibration key to lead the STD Gas Calibration screen 7 On the STD Gas Calibration Step1 s...

Страница 83: ...0 automatically 10 After completion of CO2 density calibration display leads the STD Gas Calibration Step5 screen CO2 incubator returns to the normal operation 11 On the STD Gas Calibration Step5 scre...

Страница 84: ...ast once a month 1 Remove all the inner attachments by the procedures shown on page 23 2 Clean the chamber and all the inner attachments by the procedures shown on page 22 3 Install all the inner atta...

Страница 85: ...nder empty The CO2 density does not increase when the CO2 valve is opened Err01 CO2 Gas Empty is displayed in the message display field Chamber temperature sensor error The chamber temperature sensor...

Страница 86: ...F Note Resolve the cause of the alarm in reference to page 83 85 because the alarm itself is not deactivated by pressing Buzzer key Table 6 In the cases of high limit temperature alarm or Err38 The ou...

Страница 87: ...when the outer door is open ON when the outer door is open H2O2 mist generation is cancelled Resolve time is extended Failure during H2O2 gas resolve by UV The UV lamp failed during H2O2 gas resolve b...

Страница 88: ...8 kgf cm2 G 551 psi G or lower it is a sign that there is little gas remaining Replace the cylinder soon Is the incubator operating beside the appliance that generates the electromagnetic wave The cha...

Страница 89: ...density is slow to recover when the CO2 gas pressure is normal it is possible that the HEPA filter may be clogged Contact our sales representative or agent Is there little gas remaining in the CO2 ga...

Страница 90: ...ator securely installed Set the 2 pins of the H2O2 generator in the 2 holes at the left bottom of the duct Fig 3 Is it the end of life of the H2O2 generator If the total operating time exceeds 5 000 h...

Страница 91: ...by qualified personnel only If the CO2 incubator is left where outsiders enter it may result unexpected accident for example children to become locked inside Before disposing the CO2 incubator with b...

Страница 92: ...heater jacket air jacket system Heater 295 W Humidifying system Natural evaporation with humidifying pan Temperature controller PID control system Temperature display Digital display CO2 controller P...

Страница 93: ...r RS 232C RS 485 Only for the Data acquisition system MTR 5000 user Note Refer to the updated catalog when ordering an optional component Designs and specifications are subject to change without notic...

Страница 94: ...nt temperature 23 o C setting 37 o C CO2 5 no load Chamber humidity 95 R H 5 R H Applicable environment condition Temperature 5 o C to 35 o C Humidity 80 R H max The designed performance may not be ob...

Страница 95: ...the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see bel...

Страница 96: ...Printed in Indonesia LDCL040000 0 N0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Отзывы: