19
5. Monitoring Operations
5.1. Operations Display Via LED
Certain operations modes of the turbopump and the TC 600
can be ascertained via the two integrated LEDs located on
the front panel of the TC 600.
The following operations modes are displayed:
5.2. Turbopump Temperature Management
Where impermissible motor temperatures are involved or the
temperature of the TC 600 casing is too high, the motor
current is reduced. This can lead to dipping below the set
rotation speed switchpoint and results in the turbomolecular
pump being switched off.
LED on the TC 600 glows red: Collective malfunction.
LED
Cause
green red
Glows
– power supply OK
– Function ”pumping station ON”
carried out
Flashes short
– power supply OK
(1/12s active)
– Pumping station OFF
Blinks
– Mains power supply failure
(1/2s active)
Glows
– Collective malfunction
(for example, run-up time error,
over-temperature, turbopump
or TC 600)
– Switching output 2 active (low)
Blinks
– Warning
(1/2s active)
(e.g. supply voltage short
circuit to earth, mains
power supply failure)
Differentiated malfunction and warning signals
are only possible with the use of the DCU or
HPU.
☞
PLEASE NOTE