3.10. Connecting The Remote Control Unit
Remote control options for various functions are provided with
the connection "REMOTE" on the TC 600 via the 15-pole D-Sub-
Connector. Shielded cable must be used.
Shielding is on the
plug side of the TC 600 connected to the TC casing.
The inputs 2-6 are activated by connecting them to the + 24 V
on pin 1 (active high) (please see Section 3.12. Connections
Diagram).
Pin occupancy and remote plug functions
(please see following table).
When connecting supply voltage, the turbopump
is started.
On delivery:
Pin 1, Pin 2, Pin 3 and Pin 4 are bridged in the
mating plug
.
CAUTION
1
8
9
15
Pin arrangement remote plug
Pin occupancy and remote plug functions
Pin nr.
Input open (low)
Input closed (high) on + 24V (pin 1)
1
+24 V
2
venting blocked (see Section 3.5.)
venting released (see Section 3.5.)
3
motor, turbopump off
motor, turbopump on:
the turbopump is driven, current flows through the motor coils
4
pumping station off
pumping station on:
the turbopump is driven, backing pump is started via the relay box
5
heating off
heating on:
the heating is switched on once the rotation speed switchpoint is attained and off
when the rotation speed switchpoint is unattained
optional: sealing gas valve off1)
optional: sealing gas valve on1)
5
reset:
by supplying a pulse (T < 2s) with an amplitude of 24V a malfunction
acknowledgement can be processed
6
standby off
standby on:
pump is accelerated to 66% of its nominal rotation speed
7
rotation speed setting mode off
the rotation speed can be changed by feeding a PWM signal to this pin or
via Serial Interface RS 485 (see Section 4.7. Rotation Speed Setting Mode)
8
Output (low)
Output (high)
rotation speed switchpoint not attained
rotation speed switchpoint attained; output can be loaded with 24 V/50 mA
9
Output (low)
Output (high)
Collective malfunction message
malfunction-free operations;
output can be loaded with 24 V/50 mA
10
Mass (ground)
_________
11
Contact
contact2) between pin 11 and pin 12 closed
Output 1: Switchpoint attained
if the turbopump is above the switchpoint
12
Contact
Output 1: Switchpoint attained
13
Contact
contact2) between pin 13 and pin 14 open in the event of a collective malfunction
Output 2: Collective malfunction message
14
Contact
Output 2: Collective malfunction message
15
Analog output
rotation speed proportional output voltage 0 - 10 VDC =
0 - 100 % * fend/load R
≥
10 k
Ω
1) The option must be set via the Serial Interface RS 485
(see Operating Instructions PM 0547 BN, Pumping Operating with DCU
2) The following technical data is applicable for the contacts:
Umax = 50 VDC
Imax = 1 A