Remove operating fluid reservoir
1. Wear laboratory gloves to avoid skin contact.
2. Place the turbopump on the closed high vacuum flange.
3. Unscrew the Torx screws from the closing cap.
4. Lift the closing cap off the pump bottom part.
5. Be careful with the seal; use an o-ring picker to remove it if needed.
6. Use tweezers to lever operating fluid reservoir out of bearing housing.
7. Pull the old capillary rods (8×) out of the pump base with the tweezers.
8. Clean closing cap with a clean, lint-free cloth.
– Do
not use cleaning agents
.
7.3.2 Assemble operating fluid reservoir
Required consumables
● Laboratory gloves
● Operating fluid reservoir
Required tools
● Torx screwdriver
TX 10
● Tweezers
● Screwdriver
● Calibrated torque wrench (tightening factor ≤ 1.6)
4
5
3
2
1
6
Fig. 15:
Assemble operating fluid reservoir
1 Operating fluid reservoir
4
Closing cap
2 Capillary rod (8×)
5
Seal
3 Torx screws
6
Protective cap
Assemble operating fluid reservoir
1. Insert all new capillary rods with the tweezers.
2. Insert new operating fluid reservoir into bearing housing.
– When doing so, do
not
exert pressure on the operating fluid reservoir.
3. Screw seal into pump bottom part.
4. Fit closing cap.
5. Tighten the closing cap.
– Tightening torque:
1.5 Nm
.