NOTICE
Damage to sealing surfaces from unsuitable tools
The use of unsuitable tools for removal or insertion of sealing rings damages the sealing surfaces,
causing vacuum pump leakage.
► Never use sharp, metallic tools (e.g. tweezers).
► Only remove sealing rings with an O-ring picker.
Replacing the operating fluid reservoir
Depending on its design, the turbopump’s operating fluid reservoir may feature capillary
rods.
● When ordering spare parts, make sure you use the correct pump article number and
the operating fluid reservoir.
● This information can be found on the pump rating plate.
You can find the safety data sheet in the
Pfeiffer Vacuum Download Center
Prerequisites
● Turbopump off
● Vacuum system vented to atmospheric pressure
● Electrical supply disconnected
● All cables disconnected
● All openings sealed with the original protective covers and any plugs
7.3.1 Remove operating fluid reservoir
Required consumables
● Clean, lint-free cloth
● Laboratory gloves
Required tools
● Torx screwdriver
TX 10
● Tweezers
● O-ring picker
2
3
1
5
4
6
Fig. 14:
Remove operating fluid reservoir
1 Torx screws
5
Capillary rod (8×)
2 Closing cap
6
Closing cap
3 Seal
7
Protective cap
4 Operating fluid reservoir
Maintenance
39/58