WARNING
Danger of cut injuries from unexpected start up.
The use of mating plugs of the electronic drive unit (accessories) enables the automatic run-up of the
vacuum pump as soon the power is turned on. Attaching mating plugs before or during the installa-
tion leads to the movement of parts hence the risk of cut injuries by sharp-edged in the exposed high
vacuum flange.
► Only connect mating plugs after the mechanical installation.
► Only switch on the vacuum pump immediately prior to operation.
DC out
S1
1
2
3
5
4
Fig. 13:
Connecting electronic drive unit to power supply pack
1 Power supply pack
4
Electronic drive unit TC 80
2 RS-485 interface
5
"X3" connector
3 Accessory connection
Original power supply packs (e.g. TPS 180) and connection cables with blocking diode are available for
the voltage supply to the TC 80 electronic drive unit. The connection cable with blocking diode prevents
power from being fed back into the supply after switching off the turbopump.
Connection cable type
Function
Connection cable with RS-485 interface and
bridges from TC 80 to power supply pack
● Voltage supply via power supply pack
● Automatic ramp-up with bridges on pins 2, 5, 7
● Connector to display and control unit via RS-485
Connection cable with RS-485 interface, ac-
cessory connections and blocking diode from
TC 80 to power supply pack
● Voltage supply via power supply pack
● Connector for accessory units with M8 plug
● Connector to display and control unit via RS-485
Connection cable with bridges from TC 80 to
power supply pack
● Voltage supply via power supply pack
● Automatic ramp-up with bridges on pins 2, 5, 7
Connection cable with bridges and accessory
connectors from TC 80 to power supply pack
● Voltage supply via power supply pack
● Automatic ramp-up with bridges on pins 2, 5, 7
● Connector for accessory units with M8 plug
Connecting electronic drive unit
1. Ensure the correct supply voltage.
2. Make sure that the “S1” power supply pack master switch is off prior to connection.
3. Plug the 15-pole connection cable plug into the "X3" connection on the electronic drive unit and
secure it.
4. Plug the connecting socket with the bayonet catch into the “DC out” connection on the power sup-
ply pack and lock it.
5.
When using a display and control unit:
Connect “RS-485” connector of a suitable connection
cable to display and control unit.
Installation
31/58
Содержание HIPACE 80 NEO
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original HIPACE 80 NEO Turbopump...
Страница 57: ...57 58...
Страница 58: ...PT0660 ed A Date 2109 P N PT0660BEN...