Display Problem
Possible causes
Remedy
E008
Connection, electronic drive
unit - pump faulty
● Connection to the pump
faulty
● Check connections
● Acknowledge only at rotation speed f = 0
E010
Internal device error
● Contact Pfeiffer Vacuum Service.
● Acknowledge only at rotation speed f = 0
E021
Electronic drive unit does
not detect the pump
● Contact Pfeiffer Vacuum Service.
● Acknowledge only at rotation speed f = 0
E043
Internal configuration error
● Contact Pfeiffer Vacuum Service.
E044
Excess temperature, elec-
tronics
● Faulty cooling
● Improve the cooling
● Check the operating conditions
E045
Excess temperature, motor
● Faulty cooling
● Improve the cooling
● Check the operating conditions
E046
Internal initialization error
● Contact Pfeiffer Vacuum Service.
E091
Internal device error
● Contact Pfeiffer Vacuum Service.
E092
Unknown connection panel
● Contact Pfeiffer Vacuum Service.
E093
Motor temperature evalua-
tion faulty
● Contact Pfeiffer Vacuum Service.
E094
Electronics temperature
evaluation faulty
● Contact Pfeiffer Vacuum Service.
E098
Internal communication er-
ror
● Contact Pfeiffer Vacuum Service.
E107
Final stage group error
● Contact Pfeiffer Vacuum Service.
● Acknowledge only at rotation speed f = 0
E108
Rotation speed measure-
ment faulty
● Contact Pfeiffer Vacuum Service.
● Acknowledge only at rotation speed f = 0
E109
Software not released
● Contact Pfeiffer Vacuum Service.
E110
Operating fluid evaluation
faulty
● Contact Pfeiffer Vacuum Service.
● Acknowledge only at rotation speed f = 0
E111
Operating fluid pump com-
munication error
● Contact Pfeiffer Vacuum Service.
● Acknowledge only at rotation speed f = 0
E112
Operating fluid pump group
error
● Contact Pfeiffer Vacuum Service.
● Acknowledge only at rotation speed f = 0
E114
Final stage temperature
evaluation faulty
● Contact Pfeiffer Vacuum Service.
E117
Excess temperature, pump
lower part
● Faulty cooling
● Improve the cooling
● Check the operating conditions
E118
Excess temperature, final
stage
● Faulty cooling
● Improve the cooling
● Check the operating conditions
E119
Excess temperature, bear-
ing
● Faulty cooling
● Improve the cooling
● Check the operating conditions
E143
Operating fluid pump ex-
cess temperature
● Faulty cooling
● Improve the cooling
● Check the operating conditions
● Acknowledge only at rotation speed f = 0
E777
Nominal rotation speed not
confirmed
● Nominal rotation speed
not confirmed after re-
placing the electronic
drive unit
● Confirm the nominal rotation speed with
[P:
777]
● Acknowledge only at rotation speed f = 0
F001
TMS heat-up time expired
● Internal timer for heat-
up monitoring exceeded
● Check the operating conditions
Malfunctions
44/60
Содержание HiCube 30 Eco
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original HICUBE ECO Turbo pumping stations...
Страница 59: ...59 60...