62 . Frontier IR Dual-Range Systems User's Guide
5.
Fit the window to the sample compartment.
Ensure the seal is fully seated and the key on the window lines up with the notch in the
main cover (Figure 25).
Figure 25 Window installed in sample compartment
6.
Secure the window to the sample compartment wall by tightening the screws using the
3 mm screwdriver (Figure 25).
NOTE: Do not over tighten the screws.
7.
Repeat steps 2–6 for the other sample compartment window.
8.
Close the sample compartment cover.
To store the windows:
1.
Open the spare window compartment cover (Figure 26).
Figure 26 Windows situated in the spare window compartment
Screw
Notch and key
Notch and
key
Sample
compartment
wall
Sample compartment wall
Screw
Compartment
desiccant indicator
Spare window
Screw
Screw
Содержание Frontier IR
Страница 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 Frontier IR Dual Range Systems User s Guide...
Страница 11: ...Warnings and Safety Information...
Страница 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Страница 34: ...34 Frontier IR Dual Range Systems User s Guide...
Страница 35: ...Unpacking and Installation...
Страница 45: ...Using the Spectrometer with Spectrum...
Страница 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Страница 57: ...Routine Maintenance...
Страница 78: ...78 Frontier IR Dual Range Systems User s Guide...
Страница 79: ...Advanced Maintenance...
Страница 94: ...94 Frontier IR Dual Range Systems User s Guide...
Страница 95: ...Appendices...