Initialization errors
If your instrument generates one of the following error messages, switch off the instrument,
then switch the instrument on again. Consult your PerkinElmer service representative if the
problem persists.
Error message
Action
Program Integrity Failed
The instrument’s firmware may be corrupt, so must be
reinstalled.
Instrument Configuration Failed
Suggests that one or more internal devices is failing. If
the problem persists, note any associated error
messages and consult your PerkinElmer service
representative.
Scan Failed
If the problem persists, note any associated error
messages and consult your PerkinElmer service
representative.
Laser Failed
Replace Laser
Consult your PerkinElmer service representative.
The beamsplitters may be missing, the laser fringe pre-
amplifier may have failed, or the laser itself may have
failed.
Refer to
Replacing the Laser and Power Supply
page 91.
Source Failed
Replace Source
on page 83.
Calibration Failed
Too little energy may be reaching the detector. Make
sure the sample compartment is clear.
Содержание Frontier IR
Страница 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 Frontier IR Dual Range Systems User s Guide...
Страница 11: ...Warnings and Safety Information...
Страница 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Страница 34: ...34 Frontier IR Dual Range Systems User s Guide...
Страница 35: ...Unpacking and Installation...
Страница 45: ...Using the Spectrometer with Spectrum...
Страница 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Страница 57: ...Routine Maintenance...
Страница 78: ...78 Frontier IR Dual Range Systems User s Guide...
Страница 79: ...Advanced Maintenance...
Страница 94: ...94 Frontier IR Dual Range Systems User s Guide...
Страница 95: ...Appendices...