Advanced Maintenance . 81
4.
Remove the sample compartment cover, if fitted; open the cover, press on the cover
retaining clip and then pull the cover vertically to remove it (Figure 42).
Figure 42 Removing the sample compartment cover
5.
Remove the fitted sampling accessory, by pulling the release handle towards you and
sliding the accessory towards you (Figure 43).
Figure 43 Removing the sampling accessory
6.
Undo the two locking bolts on the front of the instrument using the hexagonal wrench
supplied; insert the wrench, so that the free end is pointing horizontally away from the
instrument and then rotate the wrench 180° in the direction indicated by the arrows in
Figure 44 Undoing the locking bolts
7.
Lift the main cover from the front.
Sample cover
release clip
Sampling accessory
release handle