Preparing the System
0993-5255
3
-9
EDLs
You must use the lamp holders numbered 1 through 4 for EDLs.
1. Open the lamp compartment cover by lifting it at the front and swinging it
back.
2. If necessary, assemble the lamp by sliding the lamp driver into the lamp
sleeve for the required element. Depress the locking pin as the sleeve slides
into place and make sure that the locking pin extends through the locating
hole in the sleeve.
3. Slide the lamp carefully into the selected lamp holder until it butts up against
the end-stop inside the holder.
4. Insert the coding plug over the receptacle on the lamp holder.
5. Insert the lamp connector into the correct EDL socket at the left-hand side of
the lamp compartment. Use the receptacle that has the same number as the
lamp holder.
6. Close the lamp compartment cover.
Lamp sleeve
Lamp driver
Locking pin
Coding plug
2
3
4
5
6
7
8
Coding plug
EDL socket
Содержание AAnalyst 700
Страница 1: ...ATOMIC SPECTROMETRY User s Guide AAnalyst 700 Atomic Absorption Spectrometer Return to Document Menu...
Страница 2: ......
Страница 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Страница 8: ...Contents C 4 0993 5255...
Страница 10: ...1 2 0993 5255...
Страница 44: ...Safety Regulatory Information 1 36 0993 5255...
Страница 46: ...Spectrometer Installation 2 2 0993 5255...
Страница 68: ...Spectrometer Installation 2 24 0993 5255...
Страница 70: ...Preparing the System 3 2 0993 5255...
Страница 84: ...System Maintenance 4 2 0993 5255...
Страница 98: ...Parts and Supplies 5 2 0993 5255...
Страница 102: ...Laboratory Requirements 6 2 0993 5255...
Страница 120: ...Spectrometer System Description 7 2 0993 5255...
Страница 136: ...Translations of Warnings 8 2 0993 5255...
Страница 146: ...Translations of Warnings 8 12 0993 5255...
Страница 185: ...0993 5255 In 1 Index...
Страница 186: ...Index In 2 0993 5255...
Страница 192: ......
Страница 193: ......