Spectrometer Installation
2
-16
0993-5255
Connecting cooling sources
Installing the cooling system
PerkinElmer offers a recirculating cooling system as an option for the
spectrometer system. This cooling system includes a pressure-sensitive safety
device that switches off the pump if the internal pressure in the circulatory system
increases to above a safe value (
see ‘Cooling the furnace’
in the Graphite
Furnace User’s Guide
).
Connecting and disconnecting the water hoses
The inlet and outlet hoses are permanently attached to the cooling system. They
have automatic connectors which are marked with the direction of the water flow.
1. Place the cooling system in a convenient position
on the floor
underneath
the spectrometer. Do not place the cooling system on the same bench as the
spectrometer since vibrations can adversely influence the spectrometer.
Make sure that:
–
You can reach the on/off switch at the rear.
–
There is space at the rear and sides of the cooling system for air to
circulate freely.
–
There is space to remove the stopper and fill the system.
–
You can see the water level indicator.
2. Pass the two hoses from the cooling system to the rear of the spectrometer.
Make sure they are not strained.
3. Push the hose connectors onto the correct fittings on the spectrometer as
shown on the following page.
The connectors are spring loaded. Push them straight onto the fittings until
they click into place.
Note:
To disconnect the water hoses:
Carefully pull the knurled locking ring down slightly, to release the lock, and
pull the connector off the fitting on the spectrometer.
Содержание AAnalyst 700
Страница 1: ...ATOMIC SPECTROMETRY User s Guide AAnalyst 700 Atomic Absorption Spectrometer Return to Document Menu...
Страница 2: ......
Страница 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Страница 8: ...Contents C 4 0993 5255...
Страница 10: ...1 2 0993 5255...
Страница 44: ...Safety Regulatory Information 1 36 0993 5255...
Страница 46: ...Spectrometer Installation 2 2 0993 5255...
Страница 68: ...Spectrometer Installation 2 24 0993 5255...
Страница 70: ...Preparing the System 3 2 0993 5255...
Страница 84: ...System Maintenance 4 2 0993 5255...
Страница 98: ...Parts and Supplies 5 2 0993 5255...
Страница 102: ...Laboratory Requirements 6 2 0993 5255...
Страница 120: ...Spectrometer System Description 7 2 0993 5255...
Страница 136: ...Translations of Warnings 8 2 0993 5255...
Страница 146: ...Translations of Warnings 8 12 0993 5255...
Страница 185: ...0993 5255 In 1 Index...
Страница 186: ...Index In 2 0993 5255...
Страница 192: ......
Страница 193: ......