Preparing the System
0993-5255
3
-5
Set the outlet gauge pressure for each gas to a value just below the maximum
value (
P
max
).
Always set the same outlet gauge pressure for air and nitrous oxide.
Never
set the outlet gauge pressure to a value exceeding the maximum
pressure (
P
max
).
Always turn on the air supply since air pressure is required for the
spectrometer’s pneumatic system.
Table 3-1. Required Pressures for the Burner Gases
Gas
Outlet Gauge Pressure
kPa
bar
psig
P
min
P
max
P
min
P
max
P
min
P
max
Acetylene
(C
2
H
2
)
90
100
0.9
1.0
13.0
14.5
Air
450
500
4.5
5.0
65.0
72.5
Nitrous oxide
(N
2
O)
450
500
4.5
5.0
65.0
72.5
W5.7
Warning: Acetylene – Explosion Hazard
Acetylene can decompose explosively at pressures higher than 103 kPa/
1.03 bar/15 psig.
Always make sure that the acetylene outlet gauge pressure is below this
value.
Содержание AAnalyst 700
Страница 1: ...ATOMIC SPECTROMETRY User s Guide AAnalyst 700 Atomic Absorption Spectrometer Return to Document Menu...
Страница 2: ......
Страница 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Страница 8: ...Contents C 4 0993 5255...
Страница 10: ...1 2 0993 5255...
Страница 44: ...Safety Regulatory Information 1 36 0993 5255...
Страница 46: ...Spectrometer Installation 2 2 0993 5255...
Страница 68: ...Spectrometer Installation 2 24 0993 5255...
Страница 70: ...Preparing the System 3 2 0993 5255...
Страница 84: ...System Maintenance 4 2 0993 5255...
Страница 98: ...Parts and Supplies 5 2 0993 5255...
Страница 102: ...Laboratory Requirements 6 2 0993 5255...
Страница 120: ...Spectrometer System Description 7 2 0993 5255...
Страница 136: ...Translations of Warnings 8 2 0993 5255...
Страница 146: ...Translations of Warnings 8 12 0993 5255...
Страница 185: ...0993 5255 In 1 Index...
Страница 186: ...Index In 2 0993 5255...
Страница 192: ......
Страница 193: ......