Spectrometer Installation
0993-5255
2
-19
Connecting external cooling sources
If you intend to run the spectrometer system without the PerkinElmer cooling
system, you must connect another suitable source of cooling, such as the main
water supply or a laboratory thermostat. If you connect the spectrometer system
to the main water supply, make sure that the water pressure is as given in Table
2-1.
Figure 2-5. Connecting the Main Water Supply
A water hose kit with pressure relief valve is provided with the spectrometer
system which you must install in the water supply line to the spectrometer. This
valve opens if the water pressure to the furnace system exceeds 4 bar. This feature
eliminates the risk of damage to the furnace system through excess water
pressure.
The cooling water must conform with the requirements listed in
Flow
indicator
Pressure
relief valve
Drain hose
Water In
Water Out
Содержание AAnalyst 700
Страница 1: ...ATOMIC SPECTROMETRY User s Guide AAnalyst 700 Atomic Absorption Spectrometer Return to Document Menu...
Страница 2: ......
Страница 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Страница 8: ...Contents C 4 0993 5255...
Страница 10: ...1 2 0993 5255...
Страница 44: ...Safety Regulatory Information 1 36 0993 5255...
Страница 46: ...Spectrometer Installation 2 2 0993 5255...
Страница 68: ...Spectrometer Installation 2 24 0993 5255...
Страница 70: ...Preparing the System 3 2 0993 5255...
Страница 84: ...System Maintenance 4 2 0993 5255...
Страница 98: ...Parts and Supplies 5 2 0993 5255...
Страница 102: ...Laboratory Requirements 6 2 0993 5255...
Страница 120: ...Spectrometer System Description 7 2 0993 5255...
Страница 136: ...Translations of Warnings 8 2 0993 5255...
Страница 146: ...Translations of Warnings 8 12 0993 5255...
Страница 185: ...0993 5255 In 1 Index...
Страница 186: ...Index In 2 0993 5255...
Страница 192: ......
Страница 193: ......