Safety/Regulatory Information
0993-5255
1
-17
Air supply
An air compressor is recommended for the air supply to the burner system.
Cylinders of compressed air should only be used as an emergency or short-term
solution.
If you use cylinder air, make sure that it is compressed air and not oxygen-
enriched air (e.g. breathing air); the use of oxygen-enriched air can cause a
flashback of the flame.
Never
use oxygen as the oxidant since this can cause an explosion.
Nitrous oxide
(dinitrogen monoxide, N
2
O)
Tubing and fittings carrying nitrous oxide must be free of all oil, grease, and
organic materials.
Spontaneous combustion may occur if nitrous oxide comes into contact with
these materials.
Use a double-stage or heated regulator for nitrous oxide to prevent freezing of
the diaphragm and a loss of pressure regulation.
Do not store nitrous oxide in close proximity to flammable gases, such as
acetylene or hydrogen.
Содержание AAnalyst 700
Страница 1: ...ATOMIC SPECTROMETRY User s Guide AAnalyst 700 Atomic Absorption Spectrometer Return to Document Menu...
Страница 2: ......
Страница 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Страница 8: ...Contents C 4 0993 5255...
Страница 10: ...1 2 0993 5255...
Страница 44: ...Safety Regulatory Information 1 36 0993 5255...
Страница 46: ...Spectrometer Installation 2 2 0993 5255...
Страница 68: ...Spectrometer Installation 2 24 0993 5255...
Страница 70: ...Preparing the System 3 2 0993 5255...
Страница 84: ...System Maintenance 4 2 0993 5255...
Страница 98: ...Parts and Supplies 5 2 0993 5255...
Страница 102: ...Laboratory Requirements 6 2 0993 5255...
Страница 120: ...Spectrometer System Description 7 2 0993 5255...
Страница 136: ...Translations of Warnings 8 2 0993 5255...
Страница 146: ...Translations of Warnings 8 12 0993 5255...
Страница 185: ...0993 5255 In 1 Index...
Страница 186: ...Index In 2 0993 5255...
Страница 192: ......
Страница 193: ......