Preparing the System
3
-10
0993-5255
Removing lamps
Do not
unplug and remove lamps while they are switched on.
1. On the
Toolbar
, click on
Lamps
. The Align Lamps window appears.
2. Make sure that the lamps you want to remove are switched off:
The
On
buttons must
not
be green. If necessary, click on the respective
On
buttons.
3. Close the Align Lamps window.
4. Unplug and remove the lamps. For EDLs, remove the coding plugs as well.
Tips for installing lamps
Use only PerkinElmer lamps. You can use both hollow cathode and
electrodeless discharge lamps.
Do not touch the front window of the lamp; perspiration or other
contamination can reduce the intensity of the radiation.
If you install more than one lamp containing a particular element, for
example, a single-element copper lamp and a multi-element lamp containing
copper, the system will use the lamp in the holder with the lower number
when you perform determinations of that element. Make sure that you place
the preferred lamp in the holder with the lower number.
With EDLs, connect each lamp plug to the socket that has the same number
as the holder of the lamp.
With EDLs, make sure that you use the correct coding plug and connect it to
the receptacle on the lamp holder containing the EDL.
If you use the wrong coding plug or connect the coding plug to the wrong
receptacle, you will cause the spectrometer to malfunction and may cause
irreparable damage to the lamp.
Содержание AAnalyst 700
Страница 1: ...ATOMIC SPECTROMETRY User s Guide AAnalyst 700 Atomic Absorption Spectrometer Return to Document Menu...
Страница 2: ......
Страница 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Страница 8: ...Contents C 4 0993 5255...
Страница 10: ...1 2 0993 5255...
Страница 44: ...Safety Regulatory Information 1 36 0993 5255...
Страница 46: ...Spectrometer Installation 2 2 0993 5255...
Страница 68: ...Spectrometer Installation 2 24 0993 5255...
Страница 70: ...Preparing the System 3 2 0993 5255...
Страница 84: ...System Maintenance 4 2 0993 5255...
Страница 98: ...Parts and Supplies 5 2 0993 5255...
Страница 102: ...Laboratory Requirements 6 2 0993 5255...
Страница 120: ...Spectrometer System Description 7 2 0993 5255...
Страница 136: ...Translations of Warnings 8 2 0993 5255...
Страница 146: ...Translations of Warnings 8 12 0993 5255...
Страница 185: ...0993 5255 In 1 Index...
Страница 186: ...Index In 2 0993 5255...
Страница 192: ......
Страница 193: ......